FAST-PULSED ELECTROCHEMICAL DETECTION AT NOBLE-METAL ELECTRODES - THE FREQUENCY-DEPENDENT RESPONSE AT GOLD ELECTRODES FOR CHROMATOGRAPHICALLY SEPARATED CARBOHYDRATES

被引:21
|
作者
ROBERTS, RE [1 ]
JOHNSON, DC [1 ]
机构
[1] IOWA STATE UNIV SCI & TECHNOL,DEPT CHEM,AMES,IA 50011
关键词
PULSED ELECTROCHEMICAL DETECTION; CARBOHYDRATES; LIQUID CHROMATOGRAPHY;
D O I
10.1002/elan.1140060402
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Waveforms used in pulsed electrochemical detection (PED) have frequencies of ca. 1 Hz when optimized to give a maximum signal-to-noise ratio (S/N) for carbohydrates. However, higher frequencies are desired for applications of PED to capillary electrophoresis (CE) and microcolumn liquid chromatography (MLC), which can produce narrow elution peaks. Minimization of the time periods for oxidative cleaning and reductive reactivation in PED waveforms generated by the Dionex Pulsed Electrochemical Detector allows the increase in waveform frequency to ca. 3 Hz without change in the traditional time period of 200 ms, prescribed for integration of the anodic current. However, further increase in frequency requires a decrease in the integration period with a corresponding loss of signal strength.
引用
收藏
页码:269 / 273
页数:5
相关论文
共 4 条