QUANTIFICATION OF THE EFFECTS OF RESPIRATION AND PARALLAX ON INFERIOR VENA-CAVAL FILTER POSITION

被引:6
|
作者
WOLFER, GK [1 ]
TAYLOR, FC [1 ]
SMITH, DC [1 ]
机构
[1] LOMA LINDA UNIV,MED CTR,DEPT RADIOL,11234 ANDERSON ST,LOMA LINDA,CA 92354
关键词
VENAE-CAVAE; FILTERS;
D O I
10.1016/S1051-0443(94)71502-5
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PURPOSE: The change in inferior vena caval (IVC) filter position at follow-up relative to the filter position at implantation has been used as a criterion for evaluation and comparison of these devices. Perceived changes in filter position may be due to respiratory movement and/or changes in parallax between the initial and follow-up imaging studies. In this study the authors evaluated and attempted to quantify the effects of respiratory movement and parallax. PATIENTS AND METHODS: After placement of an IVC filter, radiographs of the abdomen were taken at maximum inspiration and maximum expiration in 30 patients. The effect of parallax on apparent filter movement was studied by using a phantom. RESULTS: The average filter movement on inspiration/expiration radiographs (corrected for magnification) was 3.6 mm +/- 2.2. An 8.5-mm maximal change secondary to parallax was seen in the phantom study. CONCLUSION: When follow-up images are obtained, efforts should be made to closely reproduce patient positioning and patient respiration to reduce errors in the interpretation of filter migration.
引用
收藏
页码:357 / 360
页数:4
相关论文
共 50 条
  • [1] INFERIOR VENA-CAVAL STENT FILTER
    WALLACE, MJ
    OGAWA, K
    WRIGHT, K
    CARRASCO, CH
    RICHLI, W
    CHARNSANGAVEJ, C
    AMERICAN JOURNAL OF ROENTGENOLOGY, 1986, 147 (06) : 1247 - 1250
  • [2] THROMBUS DEFORMATION OF INFERIOR VENA-CAVAL FILTER
    MARX, M
    MEYER, S
    MALHOTRA, U
    AMERICAN JOURNAL OF ROENTGENOLOGY, 1995, 164 (03) : 771 - 772
  • [3] THE BIRDS NEST INFERIOR VENA-CAVAL FILTER
    ROEHM, JOF
    GIANTURCO, C
    BARTH, MH
    SEMINARS IN INTERVENTIONAL RADIOLOGY, 1986, 3 (03) : 205 - 213
  • [4] SPLAYING OF TITANIUM GREENFIELD INFERIOR VENA-CAVAL FILTER
    RAMCHANDANI, P
    KOOLPE, HA
    ZEIT, RM
    AMERICAN JOURNAL OF ROENTGENOLOGY, 1990, 155 (05) : 1103 - 1104
  • [5] EARLY EXPERIENCE WITH THE GUNTHER INFERIOR VENA-CAVAL FILTER
    HUNTER, JV
    IRVING, D
    REIDY, J
    BRITISH JOURNAL OF RADIOLOGY, 1987, 60 (716): : 802 - 802
  • [6] SPONTANEOUS FRACTURE OF A VENA-TECH INFERIOR VENA-CAVAL FILTER
    AWH, MH
    TAYLOR, FC
    LU, CT
    AMERICAN JOURNAL OF ROENTGENOLOGY, 1991, 157 (01) : 177 - 178
  • [7] INFERIOR VENA-CAVAL FILTERS
    GREENFIELD, LJ
    PROCTOR, MC
    CURRENT OPINION IN CARDIOLOGY, 1994, 9 (05) : 606 - 611
  • [8] INFERIOR VENA-CAVAL FILTERS
    GUNTHER, RW
    CLINICAL RADIOLOGY, 1987, 38 (05) : 555 - 555
  • [9] TECHNICAL MODIFICATIONS IN THE PLACEMENT OF INFERIOR VENA-CAVAL FILTER DEVICES
    MENZOIAN, JO
    LOGERFO, FW
    DOYLE, JE
    WEITZMAN, AF
    SEQUEIRA, JC
    AMERICAN JOURNAL OF SURGERY, 1981, 142 (02): : 216 - 218
  • [10] INFERIOR VENA-CAVAL FILTER PLACEMENT WITHOUT PREOPERATIVE VENACAVOGRAPHY
    MIRANDE, RA
    MIRANDE, KK
    LIU, T
    BAKER, WH
    VASCULAR SURGERY, 1995, 29 (06): : 469 - 475