FOURIER-TRANSFORM RAMAN-SPECTROSCOPY WITH POLARIZATION MODULATION

被引:5
|
作者
POLAVARAPU, PL
机构
[1] Department of Chemistry, Vanderbilt University, Nashville
关键词
D O I
10.1016/0584-8539(90)80087-F
中图分类号
O433 [光谱学];
学科分类号
0703 ; 070302 ;
摘要
Fourier Transform Raman spectroscopy can be performed with either a Martin-Puplett interferometer (MPI) or a Michelson interferometer (MI). With linearly polarized exciting light, the MPI provides information on the Stokes parameters S1, and S3 of the scattered Raman light, from a single measurement. The parameter S1 represents the difference in the scattered intensities with parallel and perpendicular polarizations while S3 represents the difference in the scattered intensities with right and left circular polarizations. A procedure to extract these parameters with a Michelson interferometer is developed in this paper. A comparison of both approaches indicates that the use of the MPI is favored if the objective is to extract the Stokes parameters. Both approaches are equally favored if the objective is to measure the depolarization ratios. For measuring total Raman scattering, the use of a Michelson interferometer appears preferable. © 1990.
引用
收藏
页码:171 / 175
页数:5
相关论文
共 50 条