共 50 条
AUTOMATED FLOW-INJECTION TECHNIQUE FOR USE IN DISSOLUTION STUDIES OF SUSTAINED-RELEASE FORMULATIONS - APPLICATION TO IRON(II) FORMULATIONS
被引:12
|作者:
GEORGIOU, CA
VALSAMI, GN
MACHERAS, PE
KOUPPARIS, MA
机构:
[1] UNIV ATHENS,DEPT CHEM,ANALYT CHEM LAB,PANEPISTIMIOPOLIS,GR-15771 ATHENS,GREECE
[2] UNIV ATHENS,DEPT PHARM,BIOPHARMACEUT & PHARMACOKINET LAB,GR-15771 ATHENS,GREECE
关键词:
IRON DETERMINATION IN DRUG FORMULATIONS;
FLOW-INJECTION ANALYSIS;
FERROZINE;
DISSOLUTION TESTING AUTOMATION;
SUSTAINED-RELEASE FORMULATIONS;
D O I:
10.1016/0731-7085(93)E0015-F
中图分类号:
O65 [分析化学];
学科分类号:
070302 ;
081704 ;
摘要:
The application of flow-injection analysis (FIA) to automated dissolution studies of sustained-release formulations is described. The long-term stability of the dissolution-FIA analyser was checked during unattended operation for 42 h. The construction of multiple calibration curves with the so-called electronic dilution FIA procedure was used to extend the linear range of the determination. The computer-controlled FIA system and the principles of associated software are described and applied to dissolution studies of sustained-release formulations of iron(II) using its sensitive reaction with the colour reagent, ferrozine. The extended linear range of the determination is 1-130 ppm iron(II) and the precision (RSD) better than 3% (n = 3).