DETERMINATION OF MIXTURES OF COBALT AND IRON BY 1ST DERIVATIVE SPECTROPHOTOMETRY

被引:18
|
作者
TORAL, MI [1 ]
RICHTER, P [1 ]
SILVA, L [1 ]
SALINAS, A [1 ]
机构
[1] UNIV CHILE,FAC SCI,DEPT CHEM,POB 653,SANTIAGO,CHILE
关键词
D O I
10.1006/mchj.1993.1094
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Cobalt and iron can be determined simultaneously by first derivative spectrophotometry. The method does not require simultaneous equations to be resolved and it is based on the separation of the analytes, by liquid-liquid extraction, integrated reaction with 3-(4-phenyl-2-pyridinyl)-1,2,4-triazine, and perchlorate in dichloroethane and on the subsequent derivative spectrophotometric measurement using the graphical and the zero-crossing methods for determination of iron and cobalt, respectively. Cobalt and iron were thus determined in the range 57-1200 ng/ml and 2.0-120 ng/ml, respectively, in the presence of one another. The method has a high sensitivity and the detection limits (3σ) were calculated to be 17.1 ng/ml of cobalt and 0.6 ng/ml of iron. The relative standard deviations were in all instances less than 3.2%. The proposed method was successfully applied to the determination of both analytes in river water. © 1993 Academic Press. All rights reserved.
引用
收藏
页码:221 / 228
页数:8
相关论文
共 50 条