Validation of ion level analysis as method to assess in vitro total hip replacement wear

被引:2
|
作者
Williams, S. [1 ]
Hardaker, C. [2 ]
Leslie, I. [1 ]
Loughran, F. [2 ]
Isaac, G. [2 ]
Fisher, J. [1 ]
机构
[1] Univ Leeds, iMBE, Leeds, W Yorkshire, England
[2] DePuy Int Ltd, Leeds, W Yorkshire, England
关键词
Hip replacements; Metal on metal; Wear;
D O I
10.1179/175158308X383170
中图分类号
TB3 [工程材料学];
学科分类号
0805 ; 080502 ;
摘要
Currently in vitro hip simulator wear of metal on metal hip replacements is carried out by gravimetric means. However, this is time consuming and cannot be used to assess wear between measurement intervals. Measuring ion levels (Co and Cr) by inductively coupled plasma mass spectroscopy has been proposed as an alternative method. A standard hip simulator test was conducted using clinically available parts and wear assessed by gravimetric means; additionally, the serum lubricant was assessed for Co and Cr ion levels. A strong correlation in the ion levels and gravimetric was observed, indicating that this method would provide an appropriate alternative wear measurement technique in hip simulator testing.
引用
收藏
页码:104 / 108
页数:5
相关论文
共 50 条
  • [1] Radiostereometric analysis using clinical radiographic views: Validation measuring total hip replacement wear
    Yuan, Xunhua
    Cheung, Kimberley Lam Tin
    Howard, James L.
    Lanting, Brent A.
    Teeter, Matthew G.
    [J]. JOURNAL OF ORTHOPAEDIC RESEARCH, 2016, 34 (09) : 1521 - 1528
  • [2] In-vitro Wear Test and Computational Wear Prediction of Total Hip Replacement in China
    Yang, Shu
    Cui, Wen
    Zhang, Xiaogang
    Zhang, Yali
    Xie, Fengbao
    Ren, Xuhui
    Chen, Zhenxian
    Jin, Zhongmin
    [J]. Mocaxue Xuebao/Tribology, 2021, 41 (06): : 1004 - 1018
  • [3] Wear and performance of a tripolar total hip replacement
    Pritchett, James W.
    [J]. ANNALS OF JOINT, 2021,
  • [4] RATE OF WEAR IN TOTAL HIP-REPLACEMENT
    CHARNLEY, J
    HALLEY, DK
    [J]. CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 1975, (112) : 170 - 179
  • [5] Assessment of polyethylene wear in total hip replacement
    Devane, PA
    Horne, JG
    [J]. CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 1999, (369) : 59 - 72
  • [6] Validation of an Optical Coordinate Measuring Machine for the Measurement of Wear at the Taper Interface in Total Hip Replacement
    Cook, R. B.
    Maul, C.
    Strickland, A. M.
    [J]. MODULARITY AND TAPERS IN TOTAL JOINT REPLACEMENT DEVICES, 2015, 1591 : 362 - 378
  • [7] Finite element analysis of polyethylene wear in total hip replacement: A literature review
    Wang, Lin
    Isaac, Graham
    Wilcox, Ruth
    Jones, Alison
    Thompson, Jonathan
    [J]. PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2019, 233 (11) : 1067 - 1088
  • [8] Femoral stem wear in cemented total hip replacement
    Zhang, H-Y
    Blunt, L.
    Jiang, X-Q
    Brown, L.
    Barrans, S.
    Zhao, Y.
    [J]. PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2008, 222 (H5) : 583 - 592
  • [9] TOTAL HIP REPLACEMENT BY METHOD OF RING
    OLSSON, K
    WETZENSTEIN, H
    [J]. ACTA ORTHOPAEDICA SCANDINAVICA, 1971, 42 (05): : 452 - +
  • [10] Comparison of measures to assess outcomes in total hip replacement surgery
    Dawson, J
    Fitzpatrick, R
    Murray, D
    Carr, A
    [J]. QUALITY IN HEALTH CARE, 1996, 5 (02): : 81 - 88