METALLIC WEAR DEBRIS IN METAL-BACKED PATELLAR FAILURE

被引:33
|
作者
STULBERG, BN
MERRITT, K
BAUER, TW
机构
[1] CASE WESTERN RESERVE UNIV,DEPT BIOMED ENGN,CLEVELAND,OH 44106
[2] ST VINCENTS HOSP,CLEVELAND CTR JOINT REPLACEMENT,DEPT ORTHOPAED,CLEVELAND,OH
[3] CLEVELAND CLIN FDN,DEPT PATHOL,CLEVELAND,OH 44195
关键词
D O I
10.1002/jab.770050103
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Eight patients with failure of metal-backed patellar components were studied for evidence of intraarticular and systemic metal deposition. Seven of the eight patients had failure of a titanium-based patellar component that then articulated with a cobalt-chromium femoral component. One patient had articulation of a cobalt-chromium patellar component against a cobalt-chromium femoral component. All components failed by the same pattern of failure. Wearing of the ultra high molecular weight polyethylene was followed by displacement of the polyethylene and then metal-to-metal contact of the components. Methods for analysis of the debris included light and electron microscopy of synovial tissue and electron dispersive x-ray analysis (EDAX) of tissue. Graphite furnace atomic absorption spectroscopy (GFAAS) was done on synovial fluid and synovial tissue as well as urine and blood specimens taken at various times from implantation to patellar revision. Massive deposition of metallic debris within the knee joint was documented both histologically and by GFAAS. Titanium levels reached as high as 2776 ppm (mg/L) in the synovial fluid and as high as 917 mg/g of synovial tissue. Elevation of all the pertinent metals was seen in the tissue and body fluids. Titanium and aluminum levels were the highest. Titanium levels in the serum generally were highest near the time of implant failure. Other metal elevations were variable and could not be related to the time of failure. Synovial fluid levels of all metals and serum levels of titanium are reflective of intraarticular generation of metallic debris. This study demonstrates that in situations of failure of a device, massive intraarticular debris generation occurs and systemic transport will occur. To date no systemic consequences can be attributed to the deposition of debris and follow-up urine and blood specimens suggest that debridement and revision of the failed device results in lowering of the ion levels. (C) 1994 John Wiley and Sons, Inc.
引用
收藏
页码:9 / 16
页数:8
相关论文
共 50 条
  • [1] FAILURE MECHANISMS OF METAL-BACKED PATELLAR COMPONENTS
    STULBERG, SD
    STULBERG, BN
    HAMATI, Y
    TSAO, A
    [J]. CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 1988, (236) : 88 - 105
  • [2] EARLY FAILURE OF METAL-BACKED PATELLAR ARTHROPLASTY
    LEVI, N
    KOFOED, H
    [J]. JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1994, 76B (04): : 675 - 675
  • [3] FURTHER OBSERVATIONS ON METAL-BACKED PATELLAR COMPONENT FAILURE
    BAYLEY, JC
    SCOTT, RD
    [J]. CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 1988, (236) : 82 - 87
  • [4] Isolated All-polyethylene Patellar Revisions for Metal-backed Patellar Failure
    Garcia, Ryan M.
    Kraay, Matthew J.
    Goldberg, Victor M.
    [J]. CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2008, 466 (11) : 2784 - 2789
  • [5] FAILURE OF THE METAL-BACKED PATELLAR COMPONENT AFTER TOTAL KNEE REPLACEMENT
    BAYLEY, JC
    SCOTT, RD
    EWALD, FC
    HOLMES, GB
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1988, 70A (05): : 668 - 674
  • [6] METAL-BACKED PATELLAR COMPONENT FAILURE AND METALLIC SYNOVITIS IN TOTAL KNEE ARTHROPLASTY - A CASE-REPORT
    HALBRECHT, J
    GIORDANO, CP
    JAFFE, WL
    [J]. BULLETIN OF THE HOSPITAL FOR JOINT DISEASES ORTHOPAEDIC INSTITUTE, 1989, 49 (01): : 48 - 54
  • [7] FAILURE OF METAL-BACKED UNCEMENTED PATELLAR COMPONENTS - REPORT ON 3 CASES
    FELMET, G
    DENICOLA, U
    SPRINGORUM, HW
    [J]. ACTA ORTHOPAEDICA SCANDINAVICA, 1989, 60 (06): : 715 - 717
  • [8] EVALUATION OF THE EFFECT OF THE FEMORAL ARTICULAR SURFACE MATERIAL ON THE WEAR OF A METAL-BACKED PATELLAR COMPONENT
    MILLIANO, MT
    WHITESIDE, LA
    KAISER, AD
    ZWIRKOSKI, PA
    [J]. CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 1993, (287) : 178 - 186
  • [9] THE USE OF METAL-BACKED PATELLAR PROSTHESES IN TOTAL KNEE ARTHROPLASTY
    LASKIN, RS
    BUCKNELL, A
    [J]. CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 1990, (260) : 52 - 55
  • [10] Metal-backed patellar component failure in total knee arthroplasty presenting as a giant calf mass
    Chang, FY
    Tseng, KF
    Chen, WM
    Huang, CK
    Chen, TH
    Lo, WH