SIMULTANEOUS DETERMINATION OF THEOPHYLLINE AND CAFFEINE BY REVERSED PHASE LIQUID-CHROMATOGRAPHY USING PHENYL COLUMN

被引:7
|
作者
WONG, SHY [1 ]
MARZOUK, N [1 ]
MCHUGH, SL [1 ]
CAZES, E [1 ]
机构
[1] IBM INSTRUMENTS INC, DANBURY, CT 06810 USA
来源
JOURNAL OF LIQUID CHROMATOGRAPHY | 1985年 / 8卷 / 10期
关键词
BIOMEDICAL ENGINEERING - Patient Treatment;
D O I
10.1080/01483918508074096
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Phenyl, C-8, and C-18 columns were evaluated for the simultaneous analysis of theophylline (TH) and caffeine (CA). The phenyl column was chosen for its capability of resolving of 1,7-dimethylxanthine from theophylline. The procedure utilized protein precipitation of either 50 or 100 mu l of serum with diluted trichloroacetic acid, followed by analysis with either a manual isocratic or automated gradient elution mode. Chromatographic conditions were: column equals phenyl (4. 5 multiplied by 150 mm with a 4. 5 multiplied by 50 mm guard column), mobile phase equals phosphate or acetate/ACN, and detection wavelength equals 280 nm.