DIFFERENTIAL CONDUCTIMETRY IN FLOW-INJECTION - DETERMINATION OF AMMONIA IN KJELDAHL DIGESTS

被引:31
|
作者
ROHWEDDER, JJR [1 ]
PASQUINI, C [1 ]
机构
[1] UNIV ESTADUAL CAMPINAS,INST QUIM,CP 6154,BR-13081 CAMPINAS,SP,BRAZIL
关键词
DIFFERENTIAL CONDUCTIVITY METER; AMMONIA DETERMINATION; FLOW INJECTION; NITROGEN IN KJELDAHL DIGESTS; CARBON DIOXIDE DETERMINATION;
D O I
10.1039/an9911600841
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A differential conductivity meter has been constructed for use in flow injection (Fl) systems. The instrument employs a Wien bridge to generate a sinusoidal alternating electrical potential (2 V, 1.4 kHz) applied to a measurement bridge containing two twin conductance flow cells. The difference in the conductance between the cells is monitored. The instrument can be used to follow the small changes in conductance that occur in samples in which a high background ionic concentration is found. This facility was used for the determination of ammonia in Kjeldahl digests. The de-ionized water acceptor stream, previously used in the Fl system, was replaced by a 5 x 10(-4) mol dm-3 HCl solution and the decrease in the conductance caused by the reaction of hydrogen ions with ammonia was monitored. Linear calibration graphs were obtained in the range 0.5-25 mmol dm-3 of ammonium ion and samples can be processed at a rate of 90 h-1. The relative standard deviation for the peak heights of the Fl signals is 1 % or lower. The differential system was briefly studied for the determination of carbon dioxide by using NaOH solution as the acceptor stream.
引用
收藏
页码:841 / 845
页数:5