Dosimetric and QA aspects of Konrad inverse planning system for commissioning intensity-modulated radiation therapy

被引:2
|
作者
Deshpande, Shrikant [1 ]
Sathiyanarayanan, V. K. [1 ]
Bhangle, Janhavi [1 ]
Swamy, Kumara [1 ]
Basu, Sumit [1 ]
机构
[1] Ruby Hall Clin, Dept Radiat Oncol, 40,Sassoon Rd, Pune 411001, Maharashtra, India
关键词
Dose volume histogram; intensity-modulated radiotherapy; multileaf collimator; treatment planning system;
D O I
10.4103/0971-6203.33240
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The intensity-modulated radiation therapy (IMRT) planning is performed using the Konrad inverse treatment planning system and the delivery of the treatment by using Siemens Oncor Impression Plus linear accelerator ( step and shoot), which has been commissioned recently. The basic beam data required for commissioning the system were generate. The quality assurance of relative and absolute dose distribution was carried out before clinical implementation. The salient features of Konrad planning system, like dependence of grid size on dose volume histogram (DVH), number of intensity levels and step size in sequencer, are studied quantitatively and qualitatively. To verify whether the planned dose [from treatment planning system (TPS)] and delivered dose are the same, the absolute dose at a point is determined using CC01 ion chamber and the axial plane dose distribution is carried out using Kodak EDR2 in conjunction with OmniPro IMRT Phantom and OmniPro IMRT software from Scanditronix Wellhofer. To obtain the optimum combination in leaf sequencer module, parameters like number of intensity levels, step size are analyzed. The difference between pixel values of optimum fluence profile and the fluence profile obtained for various combinations of number of intensity levels and step size is compared and plotted. The calculations of the volume of any RT structure in the dose volume histogram are compared using grid sizes 3 mm and 4 mm. The measured and planned dose at a point showed good agreement (<3%) except for a few cases wherein the chamber was placed in a relatively high dose gradient region. The axial plane dose distribution using film dosimetry shows excellent agreement (correlation coefficient >0.97) in all the cases. In the leaf sequencer module, the combination of number of intensity level 7 with step size of 3 is the optimal solution for obtaining deliverable segments. The RT structure volume calculation is found to be more accurate with grid size of 3 mm for clinical use. Thus a study regarding various aspects of commissioning of the Konrad inverse planning system for IMRT has been presented, which has been implemented in our clinic.
引用
收藏
页码:51 / 55
页数:5
相关论文
共 50 条
  • [1] A framework for combining dosimetric and biological objectives in inverse planning for intensity-modulated radiation therapy
    Wiesmeyer, M
    [J]. RADIOTHERAPY AND ONCOLOGY, 2003, 68 : S101 - S101
  • [2] An Automated Intensity-Modulated Radiation Therapy Planning System
    Ahmed, Shabbir
    Gozbasi, Ozan
    Savelsbergh, Martin
    Crocker, Ian
    Fox, Tim
    Schreibmann, Eduard
    [J]. INFORMS JOURNAL ON COMPUTING, 2010, 22 (04) : 568 - 583
  • [3] On linear infeasibility arising in intensity-modulated radiation therapy inverse planning
    Censor, Yair
    Ben-Israel, Adi
    Xiao, Ying
    Galvin, James M.
    [J]. LINEAR ALGEBRA AND ITS APPLICATIONS, 2008, 428 (5-6) : 1406 - 1420
  • [4] Dosimetric study of beam angle optimization in intensity-modulated radiation therapy planning
    Shukla, Arvind Kumar
    Kumar, Sanjeev
    Sandhu, I. S.
    Oinam, Arun S.
    Singh, Ranjit
    Kapoor, Rakesh
    [J]. JOURNAL OF CANCER RESEARCH AND THERAPEUTICS, 2016, 12 (02) : 1045 - 1049
  • [5] In vivo dosimetric verification of intensity-modulated radiation therapy
    Chytyk-Praznik, K.
    Van Uytven, E.
    Van Beek, T.
    McCurdy, B. M. C.
    [J]. MEDICAL PHYSICS, 2012, 39 (07) : 4629 - 4629
  • [6] An inverse therapy planning system for intensity-modulated full body irradiation
    Bogner, L.
    Diermeier, J.
    Scherf, C.
    Moravek, Z.
    [J]. STRAHLENTHERAPIE UND ONKOLOGIE, 2006, 182 : 104 - 104
  • [7] Inverse planning for functional image-guided intensity-modulated radiation therapy
    Xing, L
    Cotrutz, C
    Hunjan, S
    Boyer, AL
    Adalsteinsson, E
    Spielman, D
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2002, 47 (20): : 3567 - 3578
  • [8] Planning and delivery of intensity-modulated radiation therapy
    Yu, Cedric X.
    Amies, Christopher J.
    Svatos, Michelle
    [J]. MEDICAL PHYSICS, 2008, 35 (12) : 5233 - 5241
  • [9] Alternative methods for intensity-modulated radiation therapy inverse planning and dose delivery
    Galvin, James M.
    [J]. SEMINARS IN RADIATION ONCOLOGY, 2006, 16 (04) : 218 - 223
  • [10] Intensity-modulated radiation therapy: The inverse, the converse, and the perverse
    Glatstein, E
    [J]. SEMINARS IN RADIATION ONCOLOGY, 2002, 12 (03) : 272 - 281