A CALIBRATION METHOD FOR 4-TERMINAL-PAIR HIGH-FREQUENCY RESISTANCE STANDARDS

被引:6
|
作者
SUZUKI, K
AOKI, T
YOKOI, K
机构
[1] Measurement Standards Center, YokogawaHewlett-Packard, Ltd., Hachioji-shi, Tokyo
关键词
D O I
10.1109/19.278586
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
A calibration method for four-terminal-pair (4TP) resistance standards (4TP-R-STD's) has been developed for frequencies up to about 10 MHz. The calibration is performed using an error-correction equation for a 4TP system. This is derived from analysis of an eight-port error adapter. The derivation of the uncertainties is also discussed.
引用
收藏
页码:379 / 384
页数:6