ANALYSIS OF SET SCREW AND SIDE-LOCK CONNECTOR RELIABILITY

被引:5
|
作者
TYERS, GFO
SANDERS, R
MILLS, P
CLARK, J
机构
[1] UNIV BRITISH COLUMBIA,VANCOUVER GEN HOSP,PACEMAKER CLIN,DIV CARDIOVASC & THORAC SURG,VANCOUVER V5Z 1M9,BC,CANADA
[2] INTERMED,FREEPORT,TX
来源
关键词
PACER HEADER; PACER CONNECTOR; PACER COMPLICATIONS;
D O I
10.1111/j.1540-8159.1992.tb03010.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
A retrospective review of complications with connectors and lead-to-header interfaces was performed following 649 pacing procedures between 1980 and 1990. There were 88 lead revisions (13.6%), 81 device replacements or modifications (12.5%), and 480 new implants (74%) using devices of five manufacturers. Two basic connector types were studied, one utilizing a set screw and the other using a side-lock compression fitting. The set screw makes electrical contact and mechanically secures the lead connector pin with a set screw insulated by a self-sealing grommet or an integral or separate set screw cover. The side-lock makes electrical contact with an automatic spring mechanism while the plastic lead terminal is secured in the connector block of the pacemaker by a Delrin side-lock compression fitting. Four hundred fifty-nine set screw connector devices were followed for up to 12 years with 14 complications (3.1%) whereas 82 side-lock lock connector devices were followed for up to 5 years with one complication (1.2%). The set screw and side-lock connectors were reliable over the period of follow-up. Although the complication rate appeared lower with the side-lock, follow-up was shorter and the number of implants smaller. With the leads used in this study, the side-lock proved to be a desirable feature due to simplicity, speed, safety, and ease of use. One limitation is the requirement for a precise IS-1 connector terminal diameter.
引用
收藏
页码:2000 / 2004
页数:5
相关论文
共 50 条
  • [1] Methodology for connector reliability analysis
    Choudhury, A
    Maxson, M
    Plucinski, M
    [J]. 47TH ELECTRONIC COMPONENTS & TECHNOLOGY CONFERENCE, 1997 PROCEEDINGS, 1997, : 928 - 935
  • [2] Reliability analysis of subsea connector structure
    Pang, Nan
    Jia, Peng
    Wang, Liquan
    Yun, Feihong
    Zhang, Ning
    Liu, Pu
    [J]. Harbin Gongcheng Daxue Xuebao/Journal of Harbin Engineering University, 2021, 42 (01): : 68 - 73
  • [3] The Strength Reliability Analysis of A Lock Ring
    Wang, Jiaoyang
    Cui, Weimin
    Han, Wei
    Yu, Tianxiang
    [J]. 2012 INTERNATIONAL CONFERENCE ON QUALITY, RELIABILITY, RISK, MAINTENANCE, AND SAFETY ENGINEERING (ICQR2MSE), 2012, : 173 - 175
  • [4] Reliability analysis of electromagnetic coil for aerospace electrical connector
    Zhao, Zhiru
    Zhou, Shengjun
    Yuan, Ruibin
    Zhao, Lanyan
    [J]. Proceedings of the 2nd International Conference on Reliability of Electrical Products and Electrical Contacts, 2007, : 366 - 368
  • [5] Biomechanical analysis of an interference screw and a novel twist lock screw design for bone graft fixation
    Asnis, S.
    Mullen, J.
    Asnis, P. D.
    Sgaglione, N.
    LaPorta, T.
    Grande, D. A.
    Chahine, N. O.
    [J]. CLINICAL BIOMECHANICS, 2017, 50 : 99 - 104
  • [6] Vibration Influence Analysis of the Reliability Test for an Undercarriagedoor Lock System
    Zhang, Tuo
    Cui, Weimin
    Guo, Wei
    Yu, Tianxiang
    [J]. AUTOMATIC MANUFACTURING SYSTEMS II, PTS 1 AND 2, 2012, 542-543 : 28 - 33
  • [7] Analysis and Calculation Method of Wear Reliability for A Lock Ring and Hook
    Han, Wei
    Cui, Weimin
    Li, Hangshuai
    Yu, Tianxiang
    [J]. 2012 INTERNATIONAL CONFERENCE ON QUALITY, RELIABILITY, RISK, MAINTENANCE, AND SAFETY ENGINEERING (ICQR2MSE), 2012, : 848 - 852
  • [8] Static model and reliability analysis of double nut ball screw considering screw bending
    Zhao, Jinsong
    Zhao, Chunyu
    Jiang, Zhiyuan
    Liu, Chang
    Xu, Mengtao
    Wen, Bangchun
    [J]. TRIBOLOGY INTERNATIONAL, 2023, 187
  • [9] Analysis of the Reliability of Subsea Clamp Connector Based on Multiple Response Surface Methodology
    An, Weizheng
    Wang, Yi
    Chen, Baijiang
    Duan, Menglan
    Zhang, Xiyang
    [J]. JOURNAL OF MARINE SCIENCE AND ENGINEERING, 2023, 11 (07)
  • [10] Bent or broken: analysis of set screw fracture in the TFNa implant
    Klima, Matthew