Three-dimensional optimization and sensitivity analysis of dental implant thread parameters using finite element analysis

被引:0
|
作者
Geramizadeh, Maryam [1 ]
Katoozian, Hamidreza [1 ]
Amid, Reza [2 ]
Kadkhodazadeh, Mahdi [2 ]
机构
[1] Amirkabir Univ, Dept Biomech Engn, Tehran, Iran
[2] Shahid Beheshti Univ Med Sci, Dent Sch, Res Inst Dent Sci, Dent Res Ctr, 7th Floor,Bldg 2 SBUMS,Arabi Ave,Daneshjoo Blvd, Tehran 1983963113, Iran
关键词
Dental implants; Thread design; Optimization; Biomechanics; Finite element;
D O I
10.5125/jkamos.2018.44.2.59
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Objectives: This study aimed to optimize the thread depth and pitch of a recently designed dental implant to provide uniform stress distribution by means of a response surface optimization method available in finite element (FE) software. The sensitivity of simulation to different mechanical parameters was also evaluated. Materials and Methods: A three-dimensional model of a tapered dental implant with micro-threads in the upper area and V-shaped threads in the rest of the body was modeled and analyzed using finite element analysis (FEA). An axial load of 100 N was applied to the top of the implants. The model was optimized for thread depth and pitch to determine the optimal stress distribution. In this analysis, micro-threads had 0.25 to 0.3 mm depth and 0.27 to 0.33 mm pitch, and V-shaped threads had 0.405 to 0.495 mm depth and 0.66 to 0.8 mm pitch. Results: The optimized depth and pitch were 0.307 and 0.286 mm for micro-threads and 0.405 and 0.808 mm for V-shaped threads, respectively. In this design, the most effective parameters on stress distribution were the depth and pitch of the micro-threads based on sensitivity analysis results. Conclusion: Based on the results of this study, the optimal implant design has micro-threads with 0.307 and 0.286 mm depth and pitch, respectively, in the upper area and V-shaped threads with 0.405 and 0.808 mm depth and pitch in the rest of the body. These results indicate that micro-thread parameters have a greater effect on stress and strain values.
引用
收藏
页码:59 / 65
页数:7
相关论文
共 50 条
  • [1] Biomechanical Effect of an Exposed Dental Implant's First Thread: A Three-Dimensional Finite Element Analysis Study
    Hudieb, Malik Ismail
    Wakabayashi, Noriyuki
    Abu-Hammad, Osama Abdullah
    Kasugai, Shohei
    MEDICAL SCIENCE MONITOR, 2019, 25 : 3933 - 3940
  • [2] Stress Analysis of Dental Implant Bodies by Three-Dimensional Finite Element Method
    Kageyama, Yoshiyuki
    Shimura, Mari
    Nagai, Eiichi
    Sato, Yoshinori
    Majima, Hideyuki
    Journal of Robotics and Mechatronics, 1995, 7 (02): : 169 - 171
  • [3] Selection of the implant thread pitch for optimal biomechanical properties: A three-dimensional finite element analysis
    Kong, Liang
    Zhao, Yunzhuan
    Hu, Kaijin
    Li, Dehua
    Zhou, Hongzhi
    Wu, Ziyan
    Liu, Baolin
    ADVANCES IN ENGINEERING SOFTWARE, 2009, 40 (07) : 474 - 478
  • [4] A three-dimensional finite element analysis with various implant designs
    Yoo, Dong-Ki
    Heo, Seong-Joo
    Koak, Jai-Young
    Kim, Seong-Kyun
    Lim, Young-Jun
    Kim, Sung-Hun
    Han, Chong-Hyun
    ASBM7: ADVANCED BIOMATERIALS VII, 2007, 342-343 : 885 - +
  • [5] Three-dimensional finite element analysis of platform switched implant
    Moon, Se-Young
    Lim, Young-Jun
    Kim, Myung-Joo
    Kwon, Ho-Beom
    JOURNAL OF ADVANCED PROSTHODONTICS, 2017, 9 (01): : 31 - 37
  • [6] Three-dimensional finite element stress analysis of uneven-threaded ti dental implant
    Li, Xingchao
    Dong, Fusheng
    INTERNATIONAL JOURNAL OF CLINICAL AND EXPERIMENTAL MEDICINE, 2017, 10 (01): : 307 - 315
  • [7] Stress distribution of two commercial dental implant systems: A three-dimensional finite element analysis
    Chang, Hao-Sheng
    Chen, Yi-Chin
    Hsieh, Yao-Dung
    Hsu, Ming-Lun
    JOURNAL OF DENTAL SCIENCES, 2013, 8 (03) : 261 - 271
  • [8] Stress distribution of the variable dynamic loading in the dental implant: A three-dimensional finite element analysis
    Djebbar N.
    Serier B.
    Bouiadjra B.B.
    1600, Trans Tech Publications Ltd (31): : 44 - 52
  • [9] Effect of Offset Implant Placement on the Stress Distribution Around a Dental Implant: A Three-Dimensional Finite Element Analysis
    Siadat, Hakimeh
    Hashemzadeh, Shervin
    Geramy, Allahyar
    Bassir, Seyed Hossein
    Alikhasi, Marzieh
    JOURNAL OF ORAL IMPLANTOLOGY, 2015, 41 (06) : 646 - 651
  • [10] Three-dimensional Finite Element Analysis of the Mechanical Properties of Helical Thread Connection
    YANG Guoqing
    HONG Jun
    ZHU Linbo
    LI Baotong
    XIONG Meihua
    WANG Fei
    Chinese Journal of Mechanical Engineering, 2013, 26 (03) : 564 - 572