EFFECTIVENESS OF PALLADIUM CHEMICAL MODIFICATION FOR THE DETERMINATION OF CADMIUM BY GRAPHITE-FURNACE ATOMIC-ABSORPTION SPECTROMETRY

被引:21
|
作者
SMEYERSVERBEKE, J
QING, Y
PENNINCKX, W
VANDERVOORT, F
机构
[1] Laboratory of Pharmaceutical and Biomedical Analyses, Farmaceutisch Instituut, Vrije Universiteit Brussel, B-1090 Brussels
[2] Liaoning Institute of Chemical Engineering, Liaoning
关键词
Cadmium determination; Graphite furnace atomic absorption spectrometry; Palladium chemical modification; Sodium chloride interference; Urine; milk and blood analysis;
D O I
10.1039/ja9900500393
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The effect of palladium on the cadmium absorbance signal was investigated. To reduce the background absorbance due to sodium chloride, large amounts of palladium should be avoided. A combination of 6 μg of palladium and 500 μg of ammonium nitrate allows the direct determination of cadmium in undiluted urine against aqueous standards. The detection limit (3σ) was calculated to be 0.1 μg l-1. This mixed modifier is shown also to be applicable to the analysis of milk and blood with a detection limit of 0.5 μg l-1.