共 24 条
ANALYTICAL FEASIBILITY OF GRAPHITE ROD VAPORIZATION HELIUM MICROWAVE-INDUCED PLASMA (GRV HE-MIP) FOR SOME NONMETAL DETERMINATIONS
被引:5
|作者:
ABDILLAHI, MM
机构:
[1] Research Institute, KFUPM, Dhahran 31261
关键词:
EMISSION SPECTROSCOPY INSTRUMENTATION;
ETV-MIP;
ANALYTICAL METHODS;
SPECTROSCOPIC TECHNIQUES;
D O I:
10.1366/0003702934066802
中图分类号:
TH7 [仪器、仪表];
学科分类号:
0804 ;
080401 ;
081102 ;
摘要:
Trace level quantities of some nonmetallic elements are determined by coupling graphite rod electrothermal vaporization with a helium microwave-induced plasma. Microliter liquid samples of chlorides, bromides, iodides, sulfides, and ammonium nitrogen have been introduced into a grooved graphite rod, where they were dried, ashed, and vaporized into the plasma. The analytical performance of the GRV/He-MIP was found to be an excellent excitation source for dry analyte vapors. The graphite rod vaporization device is well suited for sample introduction for a low-power microwave-induced plasma. Analytical calibration curves were obtained that have linearities between 2.5 to 5 orders of magnitude. Detection limits of the technique for 10-muL aqueous sample solutions are reported in the lower nanogram region. The common anions and cations did not show significant interference effects on the determination of traces of these nonmetals.
引用
收藏
页码:366 / 374
页数:9
相关论文