SPECTRAL CHARACTERISTICS OF EMBOLIC HEART SOUNDS DETECTED BY PRECORDIAL DOPPLER ULTRASOUND DURING VENOUS AIR-EMBOLISM IN DOGS

被引:2
|
作者
LUI, PW
LIN, YM
CHAN, FHY
TSOU, MY
WANG, S
LAM, FK
POON, PWF
机构
[1] UNIV HONG KONG, FAC MED, DEPT PHYSIOL, SASSOON RD, HONG KONG, HONG KONG
[2] NATL YANG MING MED COLL, TAIPEI, TAIWAN
[3] VET GEN HOSP, TAIPEI, TAIWAN
[4] UNIV HONG KONG, FAC MED, DEPT ELECTR ENGN, HONG KONG, HONG KONG
关键词
EMBOLISM; VENOUS AIR; MONITORING; PRECORDIAL DOPPLER ULTRASOUND;
D O I
10.1093/bja/71.5.689
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
We have studied the frequency spectrum of embolic heart sounds as detected by precordial Doppler ultrasound after injecting air 0.01-0.1 ml into the external jugular vein of anaesthetized dogs (n = 7). At volumes of 0.05 ml and greater, there was a significant increase in both the maximum frequency of the spectrum and the power of those high frequency components to greater than control values. Both changes correlated significantly with the volume of air injected (P < 0.05). With repeated injections of air at 5-min intervals, there was a sma# increase in the high frequency components in the control signal between injections (P < 0.05). We conclude that such spectral changes could be used to quantify the volume of air in transit through the heart, and to reflect the short-term history of venous air embolism.
引用
收藏
页码:689 / 695
页数:7
相关论文
共 12 条
  • [1] Wavelet analysis of embolic heart sound detected by precordial Doppler ultrasound during continuous venous air embolism in dogs
    Lui, PW
    Chan, BCB
    Chan, FHY
    Poon, PWF
    Wang, H
    Lam, FK
    ANESTHESIA AND ANALGESIA, 1998, 86 (02): : 325 - 331
  • [2] PRECORDIAL DOPPLER IS NOT OBSOLETE FOR VENOUS AIR-EMBOLISM MONITORING
    GIBBY, GL
    ANESTHESIOLOGY, 1988, 68 (05) : 829 - 829
  • [3] PRECORDIAL DOPPLER PROBE PLACEMENT FOR DETECTING VENOUS AIR-EMBOLISM REASSESSED
    CRAFT, RM
    WEGLINSKI, MR
    PERKINS, WJ
    LOSASSO, TJ
    ANESTHESIOLOGY, 1994, 81 (3A) : A230 - A230
  • [4] PRECORDIAL DOPPLER MONITORING AND PULSE OXIMETRY DURING CESAREAN DELIVERY - DETECTION OF VENOUS AIR-EMBOLISM
    VARTIKAR, JV
    JOHNSON, MD
    DATTA, S
    REGIONAL ANESTHESIA, 1989, 14 (03) : 145 - 148
  • [5] COMPARISON OF A TRANSESOPHAGEAL AND PRECORDIAL ULTRASONIC DOPPLER SENSOR IN THE DETECTION OF VENOUS AIR-EMBOLISM
    MUZZI, DA
    LOSASSO, TJ
    BLACK, S
    NISHIMURA, R
    ANESTHESIA AND ANALGESIA, 1990, 70 (01): : 103 - 104
  • [6] JUNCTIONAL RHYTHM CAN MIMIC AIR-EMBOLISM DURING PRECORDIAL DOPPLER MONITORING
    AZAD, SS
    MAGUIRE, D
    GOLDBERG, ME
    ANESTHESIA AND ANALGESIA, 1986, 65 (10): : 1083 - 1084
  • [7] PRECORDIAL DOPPLER DIAGNOSIS OF HEMODYNAMICALLY COMPROMISING AIR-EMBOLISM DURING CESAREAN-SECTION
    FONG, J
    GADALLA, F
    GIMBEL, AA
    CANADIAN JOURNAL OF ANAESTHESIA-JOURNAL CANADIEN D ANESTHESIE, 1990, 37 (02): : 262 - 264
  • [8] PRECORDIAL DOPPLER DIAGNOSIS OF HEMODYNAMICALLY COMPROMISING AIR-EMBOLISM DURING CESAREAN-SECTION
    LANG, S
    CANADIAN JOURNAL OF ANAESTHESIA-JOURNAL CANADIEN D ANESTHESIE, 1991, 38 (02): : 255 - 256
  • [9] Precordial Doppler probe placement for optimal detection of venous air embolism during craniotomy
    Schubert, A
    Deogaonkar, A
    Drummond, JC
    ANESTHESIA AND ANALGESIA, 2006, 102 (05): : 1543 - 1547
  • [10] Volume estimation by wavelet transform of Doppler heart sound during venous air embolism in dogs
    Chan, FHY
    Chan, BCB
    Lui, PW
    Poon, PWF
    Lam, FK
    PROCEEDINGS OF THE 20TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOL 20, PTS 1-6: BIOMEDICAL ENGINEERING TOWARDS THE YEAR 2000 AND BEYOND, 1998, 20 : 1463 - 1466