COMPARISON OF SERUM COPPER DETERMINATION BY COLORIMETRIC AND ATOMIC-ABSORPTION SPECTROMETRIC METHODS IN 7 DIFFERENT LABORATORIES

被引:7
|
作者
ARNAUD, J
CHAPPUIS, P
ZAWISLAK, R
HOUOT, O
JAUDON, MC
BIENVENU, F
BUREAU, F
机构
[1] UNIV GRENOBLE,F-38043 GRENOBLE 9,FRANCE
[2] HOP LARIBOISIERE,CENT BIOCHIM LAB,F-75475 PARIS 10,FRANCE
[3] FAC MED STRASBOURG,INST CHIM BIOL,ANAL LAB,F-67085 STRASBOURG,FRANCE
[4] CTR MED PREVENT,F-54500 VANDOEUVRE NANCY,FRANCE
[5] HOP LA PITIE SALPETRIERE,BIOCHIM LAB,F-75651 PARIS 13,FRANCE
[6] HOP DEBROUSSE,BIOCHIM LAB,F-69322 LYON,FRANCE
[7] CTR HOSP REG & UNIV CAEN,BIOCHIM LAB A,F-14040 CAEN,FRANCE
关键词
COPPER; ATOMIC ABSORPTION SPECTROMETRY; COLORIMETRY; INTERLABORATORY COMPARISON;
D O I
10.1016/0009-9120(93)90016-Y
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
An interlaboratory collaborative trial was conducted on the determination of serum copper using two different methods, based on colorimetry (test combination Copper, Boehringer Mannheim, Mannheim, Germany) and flame atomic absorption spectrometry (FAAS). The general performance of the colorimetric method was below that of FAAS, except for sensitivity and linear range, as assessed by detection limit (0.44 versus 1.32 mumol/L) and upper limit of linearity (150 versus 50 mumol/L). The range of the between-run CVs and the recovery of standard additions were, respectively, 2.3-11.9% and 92-127% for the colorimetric method and 1.1-6.0% and 93-101% for the FAAS method. Interferences were minimal with both methods. The two techniques correlated satisfactorily (the correlation coefficients ranged from 0.945-0.970 among laboratories) but the colorimetric assay exhibited slightly higher results than the FAAS method. Each method was transferable among laboratories.