THE EVALUATION OF A PORTABLE CLINICAL ANALYZER IN THE EMERGENCY DEPARTMENT

被引:41
|
作者
WOO, J [1 ]
MCCABE, JB [1 ]
CHAUNCEY, D [1 ]
SCHUG, T [1 ]
HENRY, JB [1 ]
机构
[1] SUNY HLTH SCI CTR,DEPT EMERGENCY MED,SYRACUSE,NY 13210
关键词
CHLORIDE; EMERGENCY DEPARTMENT TESTING; GLUCOSE; HEMATOCRIT; PORTABLE CLINICAL ANALYZER; POTASSIUM; RAPID REPORTING TIME; SODIUM; UREA NITROGEN;
D O I
10.1093/ajcp/100.6.599
中图分类号
R36 [病理学];
学科分类号
100104 ;
摘要
A recently available portable clinical analyzer (PCA), which examines sodium, potassium, chloride, glucose, urea nitrogen, and hematocrit levels on 60 muL of blood and calculates hemoglobin and osmolality levels within 2 minutes, was evaluated. Blood from 574 patients was drawn by emergency department staff, who immediately tested the samples with the PCA and transported them for plasma analysis on a reference analyzer in the clinical laboratory. Correlations between the PCA and the reference analyzer were as follows: R2 = 0.987 for urea nitrogen; R2 = 0.97, glucose; R2 = 0.937, K: R2 = 0.79, hematocrit; R2 0.751, sodium; and R2 = 0.689 for chloride. With its rapid turnaround, small sample requirement, and ease of operation, the PCA is most useful in an emergency department setting, where immediate access to clinically relevant laboratory testing is required in support of urgent clinical decision-making.
引用
收藏
页码:599 / 605
页数:7
相关论文
共 50 条
  • [1] EVALUATION OF A PORTABLE CLINICAL BLOOD ANALYZER IN THE EMERGENCY DEPARTMENT
    SANDS, VM
    AUERBACH, PS
    BIRNBAUM, J
    GREEN, M
    ACADEMIC EMERGENCY MEDICINE, 1995, 2 (03) : 172 - 178
  • [2] Evaluation of a portable clinical Analyzer in the pediatric emergency department: Analysis of cost and turnaround time
    Mainor, BH
    Hardwick, WE
    King, WD
    SOUTHERN MEDICAL JOURNAL, 2002, 95 (06) : 634 - 638
  • [3] Evaluation of a portable clinical analyzer in a veterinary hospital setting
    Grosenbaugh, DA
    Gadawski, JE
    Muir, WW
    JOURNAL OF THE AMERICAN VETERINARY MEDICAL ASSOCIATION, 1998, 213 (05) : 691 - 694
  • [4] Evaluation of the i-STAT Portable Clinical Analyzer in dogs
    Matburger, C
    Henke, J
    Hirschberger, J
    Matburger, S
    Erhardt, W
    TIERARZTLICHE PRAXIS AUSGABE KLEINTIERE HEIMTIERE, 2000, 28 (02): : 132 - 137
  • [5] Evaluation of a portable clinical analyzer in cynomolgus macaques (Macaca fasicularis)
    Hopper, Kelly J.
    Cray, Carolyn
    JOURNAL OF THE AMERICAN ASSOCIATION FOR LABORATORY ANIMAL SCIENCE, 2007, 46 (03): : 53 - 57
  • [6] The value of portable ultrasound for evaluation of cardiomegaly patients presenting at the emergency department
    Liu, SC
    Chang, WT
    Huang, CH
    Weng, TI
    Matthew, HMM
    Chen, WJ
    RESUSCITATION, 2005, 64 (03) : 327 - 331
  • [7] Evaluation of the i-STAT portable clinical analyzer for use in a helicopter
    Burritt, MF
    Santrach, PJ
    Hankins, DG
    Herr, D
    Newton, NC
    SCANDINAVIAN JOURNAL OF CLINICAL & LABORATORY INVESTIGATION, 1996, 56 : 121 - 128
  • [8] Evaluation of i-STAT portable clinical analyzer in a hemodialysis unit
    Gault, MH
    Harding, CE
    CLINICAL BIOCHEMISTRY, 1996, 29 (02) : 117 - 124
  • [9] EVALUATION OF IMI PORTABLE OXYGEN ANALYZER
    CHRISTIE, L
    KNIGHT, H
    MILLER, J
    LEIGH, JM
    PRESTON, TD
    ANAESTHESIA, 1975, 30 (04) : 511 - 513
  • [10] CLINICAL GUIDELINES FOR THE EMERGENCY DEPARTMENT EVALUATION OF SUBARACHNOID HEMORRHAGE
    Meurer, William J.
    Walsh, Brian
    Vilke, Gary M.
    Coyne, Christopher J.
    JOURNAL OF EMERGENCY MEDICINE, 2016, 50 (04): : 696 - 701