Static and mobile DXA scanner in-vivo cross-calibration study

被引:1
|
作者
Griffin, J. G. L. [1 ]
Knapp, K. M. [2 ]
Pearce, G. [1 ]
机构
[1] Plymouth Hosp NHS Trust, Derriford Hosp, Healthy Bones Unit, Plymouth PL6 8DH, Devon, England
[2] Univ Exeter, Exeter, Devon, England
关键词
DXA; Cross-calibration; Bone mineral density; Quality assurance; Clinical practice; Osteoporosis;
D O I
10.1016/j.radi.2012.08.005
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
In-vitro cross-calibration of DXA scanning equipment, with a phantom device, has been recommended for assessing agreement between devices co-located within DXA scanning services. This study evaluated in-vivo and in-vitro cross-calibration of a static and a mobile DXA scanner within the same service in their individual clinical settings. 50 individuals from a volunteer group were recruited to take part in this study and had DXA measurements made on two GE Lunar Prodigy Advance (GE Lunar, Bedford, UK) scanners. Results in this study showed that the scanners agreed, with no clinically significant differences in BMD measurements made at the same site on the individual devices used in this study. The in-vivo crosscalibration of the instruments was a useful experience, which demonstrated closely calibrated systems and raised the profile of the bone densitometry service within the hospital. (C) 2012 The College of Radiographers. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:7 / 10
页数:4
相关论文
共 50 条
  • [1] In-vivo cross-calibration of a new peripheral DXA densitometer.
    Patel, R
    Blake, GM
    Fogelman, I
    [J]. JOURNAL OF BONE AND MINERAL RESEARCH, 2001, 16 : S338 - S338
  • [2] CROSS-CALIBRATION OF DXA SCANNERS FOR SPINE MEASUREMENTS
    TOTHILL, P
    [J]. OSTEOPOROSIS INTERNATIONAL, 1995, 5 (05) : 410 - 411
  • [3] A Comparison of Phantoms for Cross-Calibration of Lumbar Spine DXA
    D. Pearson
    S. A. Cawte
    D. J. Green
    [J]. Osteoporosis International, 2002, 13 : 948 - 954
  • [4] Comparison of phantoms for cross-calibration of DXA scanners.
    Mathur, AK
    Pekrul, A
    Fuerst, TP
    Lu, Y
    Gcnant, HK
    [J]. JOURNAL OF BONE AND MINERAL RESEARCH, 1996, 11 : S605 - S605
  • [5] A comparison of phantoms for cross-calibration of lumbar spine DXA
    Pearson, D
    Cawte, SA
    Green, DJ
    [J]. OSTEOPOROSIS INTERNATIONAL, 2002, 13 (12) : 948 - 954
  • [6] Importance of Cross-Calibration When Replacing DXA Scanners
    Berry, Jean Kathleen
    Covey, Margaret Katherine
    Hacker, Eileen Danaher
    Smith, Donald Lawrence
    [J]. OBESITY, 2009, 17 : S204 - S204
  • [7] VALUE AND LIMITATIONS OF CROSS-CALIBRATION IN A MULTICENTER STUDY INVOLVING DIFFERENT DXA BRANDS
    SLOSMAN, DO
    DELMAS, PD
    SEBERT, JL
    REGINSTER, JY
    DUBOEUF, F
    BIRMAN, P
    TSOUDEROS, Y
    MEUNIER, PJ
    [J]. JOURNAL OF BONE AND MINERAL RESEARCH, 1993, 8 : S348 - S348
  • [8] Replacing DXA scanners: Cross-calibration with phantoms may be misleading
    Blake, GM
    [J]. CALCIFIED TISSUE INTERNATIONAL, 1996, 59 (01) : 1 - 5
  • [9] SPINE PHANTOMS ARE INADEQUATE FOR DXA WHOLE BODY COMPOSITION CROSS-CALIBRATION
    Krueger, D.
    Libber, J.
    Sanfillippo, J.
    Yu, H. J.
    Miller, C.
    Binkley, N.
    [J]. OSTEOPOROSIS INTERNATIONAL, 2014, 25 : S349 - S349
  • [10] In-vivo cross-calibration of a new peripheral densitometer: The Osteometer Dexacare G4.
    Patel, R
    Blake, GM
    Fogelman, I
    [J]. JOURNAL OF BONE AND MINERAL RESEARCH, 2003, 18 : S313 - S313