MEASUREMENT OF POLYETHYLENE WEAR IN METAL-BACKED ACETABULAR CUPS .2. CLINICAL-APPLICATION

被引:0
|
作者
DEVANE, PA [1 ]
BOURNE, RB [1 ]
RORABECK, CH [1 ]
MACDONALD, S [1 ]
ROBINSON, EJ [1 ]
机构
[1] UNIV WESTERN ONTARIO HOSP,LONDON,ON N6A 5A5,CANADA
关键词
D O I
暂无
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
The new 3-dimensional technique was applied to the radiographs of 141 patients who had received Porous Coated Anatomic total hip prostheses, Values were obtained for the position of the acetabular cup, the 3-dimensional distance and direction of femoral head displacement, and the minimum volume of polyethylene debris produced. Mean age of the patients at the time of replacement was 61 years old; mean followup was 5.6 years (range, 4-7.2 years). Change in the position of the femoral head between initial and long-term followup films was assumed to represent polyethylene wear, and formulas were used to calculate the minimal volume of polyethylene debris produced, The overall rate of 3-dimensional femoral head displacement was 0.264 mm per year, almost twice that usually quoted in the current litera-ture for 2-dimensional linear wear. A significant contribution to this value was made by anterior and posterior displacement. Two-dimensional femoral head displacement (measured in the plane of anteroposterior radiographs) on the same patients was 0.149 mm per year. The mean minimum volume of polyethylene debris produced after 5.6 years was 0.448 cm(3) (range, 0.00-2.83 cm(3)), giving a mean rate of 0.079 cm(3) polyethylene debris produced each year. Thirteen patients in this series had radiologic osteolysis and a significantly greater femoral head displacement and polyethylene volumetric wear than those with no osteolysis. Patients with a 32-mm femoral head diameter and a polyethylene liner <1 cm thick had a significantly greater amount of polyethylene wear. This series gives previously unavailable data on 3-dimensional femoral head displacement and is the first report that correlates the minimum volume of polyethylene wear produced with radiologic osteolysis.
引用
收藏
页码:317 / 326
页数:10
相关论文
共 50 条
  • [1] POLYETHYLENE WEAR IN CEMENTED METAL-BACKED ACETABULAR CUPS
    CATES, HE
    FARIS, PM
    KEATING, EM
    RITTER, MA
    [J]. JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1993, 75 (02): : 249 - 253
  • [2] MEASUREMENT OF POLYETHYLENE WEAR IN METAL-BACKED ACETABULAR CUPS .1. 3-DIMENSIONAL TECHNIQUE
    DEVANE, PA
    BOURNE, RB
    RORABECK, CH
    HARDIE, RM
    HORNE, JG
    [J]. CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 1995, (319) : 303 - 316
  • [3] Fixation of ultrahigh-molecular-weight polyethylene liners to metal-backed acetabular cups
    Williams, VG
    Whiteside, LA
    White, SE
    McCarthy, DS
    [J]. JOURNAL OF ARTHROPLASTY, 1997, 12 (01): : 25 - 31
  • [4] Polyethylene wear and periprosthetic osteolysis in metal-backed acetabular components with cylindrical liners
    Perez, RE
    Rodriguez, JA
    Deshmukh, RG
    Ranawat, CS
    [J]. JOURNAL OF ARTHROPLASTY, 1998, 13 (01): : 1 - 7
  • [5] METAL-BACKED ACETABULAR CUPS IN TOTAL HIP-ARTHROPLASTY
    RITTER, MA
    KEATING, EM
    FARIS, PM
    BRUGO, G
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1990, 72A (05): : 672 - 677
  • [6] A new approach for assessment of wear in metal-backed acetabular cups using computed tomography - A phantom study with retrievals
    Jedenmalm, Anneli
    Noz, Marilyn E.
    Olivecrona, Henrik
    Olivecrona, Lofta
    Stark, Andre
    [J]. ACTA ORTHOPAEDICA, 2008, 79 (02) : 218 - 224
  • [7] Dissociation of a polyethylene liner in a nonmodular, cemented, metal-backed acetabular component
    Bear, BJ
    Laskin, R
    Higgins, L
    [J]. ORTHOPEDICS, 1997, 20 (12) : 1179 - 1180
  • [8] ACETABULAR PRESS-FIT, METAL-BACKED CUPS. RESULTS AT FIVE YEARS
    Bistolfi, A.
    Rosso, F.
    Colzani, G.
    Belsanti, S. V.
    Massazza, G.
    Crova, M.
    [J]. MINERVA ORTOPEDICA E TRAUMATOLOGICA, 2009, 60 (01) : 1 - 7
  • [9] PELVIC STRESSES INVITRO .2. A STUDY OF THE EFFICACY OF METAL-BACKED ACETABULAR PROSTHESES
    FINLAY, JB
    BOURNE, RB
    LANDSBERG, RPD
    ANDREAE, P
    [J]. JOURNAL OF BIOMECHANICS, 1986, 19 (09) : 715 - 725
  • [10] Wear-through of metal-backed acetabular shell in total hip arthroplasty
    Huang, KY
    Liu, HC
    Hou, SM
    [J]. JOURNAL OF THE FORMOSAN MEDICAL ASSOCIATION, 2003, 102 (05) : 334 - 337