MEASUREMENT AND PREDICTION OF DRUG SOLVENT LOSSES BY EVAPORATION FROM A JET NEBULIZER

被引:11
|
作者
SMYE, SW
JOLLIE, MI
CUNLIFFE, H
LITTLEWOOD, JM
机构
[1] ST JAMES UNIV HOSP,DEPT PHARM,LEEDS LS9 7TF,W YORKSHIRE,ENGLAND
[2] ST JAMES UNIV HOSP,REG CYST FIBROSIS UNIT,LEEDS LS9 7TF,W YORKSHIRE,ENGLAND
来源
关键词
D O I
10.1088/0143-0815/13/2/004
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
The evaporative losses of solvent (water) from a commercial jet nebuliser (Unineb, Unimed Ltd) were estimated and found to be approximately 14% of total nebuliser output. A simple analysis of the nebulisation process then enabled the amount of the drug colomycin released as potentially respirable aerosol to be predicted from measurements of the total nebulisation time. Good agreement was found between the predicted (P) and measured (M) values such that P = 1.07 M (r2 = 0.98). The analysis also indicates that the proportion of the drug released as aerosol will increase as nebulisation proceeds provided nebuliser output is continuous and that for a given initial mass of drug a larger amount of drug is released as an aerosol if the volume of administration used in the nebuliser is as large as possible.