THE AUTOMATIC IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR - EFFECT OF PATCH POLARITY ON DEFIBRILLATION THRESHOLD

被引:40
|
作者
ONEILL, PG
BOAHENE, KA
LAWRIE, GM
HARVILL, LF
PACIFICO, A
机构
[1] BAYLOR UNIV,METHODIST HOSP,CARDIAC ELECTROPHYSIOL UNIT,6550 FANNIN,SM-1246,HOUSTON,TX 77030
[2] BAYLOR UNIV,DEPT MED CARDIOL,CARDIAC ELECTROPHYSIOL UNIT,HOUSTON,TX 77030
[3] BAYLOR UNIV,DEPT SURG,HOUSTON,TX 77030
关键词
D O I
10.1016/S0735-1097(10)80187-1
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
An automatic implantable cardioverter-defibrillator (AICD) was implanted in 40 patients with sudden cardiac arrest (n = 29), sustained monomorphic ventricular tachycardia (n = 10) or recurrent syncope (n = 1) who were unsuitable for direct ablative surgery or had had unsuccessful medical therapy. The effect of patch electrode polarity on the defibrillation threshold was prospectively evaluated. Two large epicardial patches were used. Initial polarity was selected at random. Ventricular fibrillation was induced by direct current and a preestablished defibrillation protocol employed to assess the minimal energy that would reproducibly defibrillate the heart. Nineteen patients had a lower defibrillation threshold with the inferior left ventricular patch as an anode and nine patients had a lower defibrillation threshold with this patch as a cathode. In general, the defibrillation threshold was lower when this patch was used as an anode than when it was used as a cathode (18 +/- 10 versus 22.6 +/- 12.2 J; p < 0.01). No preoperative variable predicted optimal polarity. Therefore, the effect of patch polarity on defibrillation threshold should be assessed in each patient at the time of AICD implantation so that the safty margin for satisfactory device function can be maximized.
引用
收藏
页码:707 / 711
页数:5
相关论文
共 50 条
  • [2] Effect of Substance Abuse on Defibrillation Threshold in Patients with Implantable Cardioverter-Defibrillator
    Perrine, Shane A.
    Nayak, Raghavendra
    Bharadwaj, Aditya S.
    Mckelvey, George
    Mohamad, Tamam
    Jacob, Sony
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2011, 34 (02): : 193 - 199
  • [3] ECHOCARDIOGRAPHIC LEFT-VENTRICULAR MASS AND DEFIBRILLATION THRESHOLD FOR THE AUTOMATIC IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR
    CHAPMAN, PD
    SAGAR, KB
    WETHERBEE, JN
    TROUP, PJ
    CIRCULATION, 1985, 72 (04) : 383 - 383
  • [4] THE AUTOMATIC IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR
    DELUCA, SA
    GIRARD, M
    AMERICAN FAMILY PHYSICIAN, 1989, 39 (01) : 141 - 143
  • [5] THE AUTOMATIC IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR
    MIROWSKI, M
    REID, PR
    MOWER, MM
    WATKINS, L
    PLATIA, EV
    GRIFFITH, LSC
    JUANTEGUY, JM
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1984, 7 (03): : 534 - 540
  • [6] AUTOMATIC IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR
    LATTUCA, J
    VELTRI, EP
    AMERICAN HEART JOURNAL, 1989, 118 (06) : 1359 - 1359
  • [7] AUTOMATIC IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR
    EPSTEIN, AE
    JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 1991, 102 (01): : 160 - 161
  • [8] Defibrillation threshold testing: Is it necessary during implantable cardioverter-defibrillator implantation?
    Liu, Qi-Ming
    Bai, Zhong-Le
    Liu, Zhen-Jiang
    Li, Xu-Ping
    Zhou, Sheng-Hua
    MEDICAL HYPOTHESES, 2009, 72 (02) : 147 - 149
  • [9] Defibrillation threshold decrease with the supradiaphragmatic extracardiac implantable cardioverter-defibrillator implantation technique
    Berruezo, Antonio
    Zeljko, Hrvojka M.
    Bartrons, Joaquim
    Mayol, Javier
    Prada, Fredy
    Brugada, Josep
    EUROPACE, 2010, 12 (11): : 1649 - 1651
  • [10] Defibrillation threshold testing: Is it really necessary at the time of implantable cardioverter-defibrillator insertion?
    Russo, AM
    Sauer, W
    Gerstenfeld, EP
    Hsia, HH
    Lin, D
    Cooper, JM
    Dixit, S
    Verdino, RJ
    Nayak, HM
    Callans, DJ
    Patel, V
    Marchlinski, FE
    HEART RHYTHM, 2005, 2 (05) : 456 - 461