COLUMN LIQUID CHROMATOGRAPHY;
NEW STATIONARY PHASES;
PEPTIDES;
PERFORMANCE TESTING;
D O I:
10.1007/BF02269742
中图分类号:
Q5 [生物化学];
学科分类号:
071010 ;
081704 ;
摘要:
Some new HPLC-phases (Polyencap, Hypercarb, Poly F, PRP-infinity) have been assessed for their suitability in the separation of peptides. Test criteria were peptide recovery, separation-efficiency, selectivity and stability. The eluents TFA/acetonitrile and acetate-buffer/propanol are compared. Of interest are the stability and recovery of Polyencap, the high separation efficiency of the standard RP-8 silica-based phase (Nucleosil C8) and the unusual peak-width of PRP-infinity. Acceptable conditions for peptide separations may be provided by Polyencap, but stability still has to be improved.