VALIDATION OF A DOPPLER GUIDE WIRE FOR INTRAVASCULAR MEASUREMENT OF CORONARY-ARTERY FLOW VELOCITY

被引:887
|
作者
DOUCETTE, JW
CORL, PD
PAYNE, HM
FLYNN, AE
GOTO, M
NASSI, M
SEGAL, J
机构
[1] UNIV CALIF SAN FRANCISCO,CARDIOVASC RES INST,SAN FRANCISCO,CA 94143
[2] SAN FRANCISCO GEN HOSP,DIV CARDIOL,SAN FRANCISCO,CA 94110
关键词
CORONARY CIRCULATION; ULTRASOUND; ECHOCARDIOGRAPHY; DOPPLER; CATHETERS;
D O I
10.1161/01.CIR.85.5.1899
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background. An improved intravascular ultrasonic Doppler device could aid the clinical assessment of coronary hemodynamics. We evaluated a new device consisting of a 12-MHz piezoelectric transducer integrated onto the tip of a 0.018-in. flexible, steerable angioplasty guide wire. Methods and Results. Doppler spectra were recorded in model tubes with pulsatile blood flow and in-line electromagnetic flowmeter. In four straight tubes (i.d., 0.79-4.76 mm), the time average of spectral peak velocity (APV) was linearly related to blood flow (Q(EMF)) (r2 greater-than-or-equal-to 0.98 for each tube). A Doppler-derived quantitative flow estimate (Q(D)) was calculated as the product of vessel cross-sectional area and mean velocity, with mean velocity estimated as 0.5 x APV. The slope of Q(D) versus Q(EMF) for the four tubes was near unity. APV was less accurate in a 7.94-mm straight tube and in tortuous segments. In four dogs, the left circumflex coronary artery (LCx) was perfused from the femoral artery via a cannula with in-line electromagnetic flowmeter. Good-quality signals were obtained in proximal and distal LCx vessels 3.3-1.2 mm in diameter. APV varied linearly with Q(EMF) (r2 greater-than-or-equal-to 0.99 in the cannula, r2 = 0.93-0.99 in proximal LCx, and r2 = 0.86-0.99 in distal LCx). Q(D) was calculated by quantitative angiography to determine proximal LCx diameter. For all dogs combined, the slope of Q(D) versus Q(EMF) was 0.95 in the cannula and 0.85 in the proximal LCx. Conclusions. The Doppler guide wire measures phasic flow velocity patterns and linearly tracks changes in flow rate in small, straight coronary arteries. It should facilitate measurement of phasic coronary flow velocity during coronary angiography and angioplasty.
引用
收藏
页码:1899 / 1911
页数:13
相关论文
共 50 条
  • [1] EVALUATION OF CORONARY-ARTERY BLOOD-FLOW BY MEASUREMENT OF CORONARY FLOW VELOCITY WITH CATHETER DOPPLER FLOWMETER
    MATSUURA, Y
    HAYASHI, T
    YASUI, K
    KURIHARA, T
    CHIN, W
    ASAO, M
    FUKUSHIMA, M
    IMAIZUMI, M
    ABE, H
    JAPANESE CIRCULATION JOURNAL-ENGLISH EDITION, 1988, 52 (09): : 999 - 999
  • [2] Comparison of intracoronary Doppler guide wire and transesophageal echocardiography in measurement of flow velocity and coronary flow reserve in the left anterior descending coronary artery
    Gadallah, S
    Thaker, KB
    Kawanishi, D
    Mehra, A
    Lau, S
    Rashtian, M
    Chandraratna, PAN
    AMERICAN HEART JOURNAL, 1998, 135 (01) : 38 - 42
  • [3] COMPARISON OF FLOW VELOCITY IN THE LEFT ANTERIOR DESCENDING CORONARY-ARTERY AND CORONARY FLOW RESERVE, BY INTRACORONARY DOPPLER FLOW WIRE AND TRANSESOPHAGEAL ECHOCARDIOGRAPHY
    THAKER, KB
    KAWANISHI, D
    NARANG, YK
    RASHTIAN, M
    MEHRA, A
    ANTHONY, P
    CHANDRARATNA, N
    CIRCULATION, 1994, 90 (04) : 276 - 276
  • [4] Coronary flow velocity changes after intravenous dipyridamole infusion: Measurements using intravascular Doppler guide wire - A documentation of flow inhomogeneity
    Voudris, V
    Manginas, A
    Vassilikos, V
    Koutelou, M
    Kantzis, J
    Cokkinos, DV
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1996, 27 (05) : 1148 - 1155
  • [5] ACCURACY OF 20-MHZ DOPPLER CATHETER CORONARY-ARTERY VELOCIMETRY FOR MEASUREMENT OF CORONARY BLOOD-FLOW VELOCITY
    TADAOKA, S
    KAGIYAMA, M
    HIRAMATSU, O
    OGASAWARA, Y
    TSUJIOKA, K
    WADA, Y
    SAWAYAMA, T
    KAJIYA, F
    CATHETERIZATION AND CARDIOVASCULAR DIAGNOSIS, 1990, 19 (03): : 205 - 213
  • [6] MEASUREMENT OF BLOOD VELOCITY IN CORONARY-ARTERY BY THE LASER DOPPLER VELOCIMETRY WITH AN OPTICAL FIBER
    TOMONAGA, G
    HOKI, N
    KAJIYA, F
    JAPANESE CIRCULATION JOURNAL-ENGLISH EDITION, 1980, 44 (08): : 631 - 632
  • [7] INTRAVASCULAR BLOOD VELOCITY IN SIMULATED CORONARY-ARTERY STENOSES
    KILPATRICK, D
    WEBBER, SB
    CATHETERIZATION AND CARDIOVASCULAR DIAGNOSIS, 1986, 12 (05): : 317 - 323
  • [8] MEASUREMENT OF CORONARY AND PERIPHERAL ARTERY FLOW BY INTRAVASCULAR ULTRASOUND AND PULSED DOPPLER VELOCIMETRY
    EICHHORN, EJ
    ALVAREZ, LG
    JESSEN, ME
    FASS, SM
    CHAO, RY
    HAAGEN, D
    GRAYBURN, PA
    AMERICAN JOURNAL OF CARDIOLOGY, 1992, 70 (04): : 542 - 545
  • [9] Validation of high temporal resolution spiral phase velocity mapping of coronary artery blood flow against Doppler flow wire
    Jennifer Keegan
    Claire E Raphael
    Robin Simpson
    Kim H Parker
    Ranil de Silva
    Carlo Di Mario
    Sanjay K Prasad
    David Firmin
    Journal of Cardiovascular Magnetic Resonance, 17 (Suppl 1)
  • [10] Coronary flow reserve measurement by guide wire-based thermodilution: Comparison to the Doppler wire
    Pijls, NHJ
    De Bruyne, B
    Bech, GJW
    van de Vosse, F
    Billet, F
    Kusters, H
    van Gelder, LM
    Nelis, O
    Bartunek, J
    Bonnier, JJRM
    Koolen, JJ
    EUROPEAN HEART JOURNAL, 2000, 21 : 34 - 34