EFFECTS OF ORGANIC MOBILE-PHASE MODIFIERS ON ELUTION AND SEPARATION OF BETA-BLOCKERS IN MICELLAR ELECTROKINETIC CAPILLARY CHROMATOGRAPHY

被引:42
|
作者
LUKKARI, P
VUORELA, H
RIEKKOLA, ML
机构
[1] UNIV HELSINKI,DEPT CHEM,DIV ANALYT CHEM,SF-00014 HELSINKI,FINLAND
[2] UNIV HELSINKI,DEPT PHARM,DIV PHARMACOGNOSY,SF-00014 HELSINKI,FINLAND
关键词
D O I
10.1016/0021-9673(93)83238-N
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A study was made of the effect of organic modifiers (acetone, acetonitrile, ethanol, ethylene glycol, methanol and 2-propanol) in phosphate buffer (0.08 M) containing 15 mM cetyltrimethylammonium bromide as surfactant on the elution and separation of eleven common beta-adrenergic blocking agents. The amount of the modifier was varied from 0.1 to 10.0% (v/v). At maximum addition, the organic solvents increased the viscosity of the buffer solution as follows: acetone 16%, acetonitrile 9%, ethanol 26%, ethylene glycol 27%, methanol 20% and 2-propanol 29%. In contrast to the migration time of the other beta-blockers, that of labetalol was not increased by the addition of organic solvent to the buffer solution. Rather, labetalol eluted more quickly with increase in the amount of modifier, and thereby effected changes in the elution order of the beta-blockers. The addition of modifiers also affected the resolution, and the best resolution values were achieved with the following amounts of organic solvent in MEKC buffer: acetone 0.1%, acetonitrile 0.1-0.5%, ethanol 5.0-7.5%, ethylene glycol 1.0-2.5%, methanol 5.0% and.-propanol 1.0-2.5% (v/v). No significant relationship was found between the elution order and separation and the structure of the beta-blockers in micellar electrokinetic capillary chromatography with an organic modifier in buffer solutions.
引用
收藏
页码:317 / 324
页数:8
相关论文
共 50 条
  • [1] EFFECT OF BUFFER SOLUTION PH ON THE ELUTION AND SEPARATION OF BETA-BLOCKERS BY MICELLAR ELECTROKINETIC CAPILLARY CHROMATOGRAPHY
    LUKKARI, P
    VUORELA, H
    RIEKKOLA, ML
    JOURNAL OF CHROMATOGRAPHY A, 1993, 652 (02) : 451 - 457
  • [2] EFFECTS OF ORGANIC MOBILE PHASE MODIFIERS IN MICELLAR ELECTROKINETIC CAPILLARY CHROMATOGRAPHY
    GORSE, J
    BALCHUNAS, AT
    SWAILE, DF
    SEPANIAK, MJ
    JOURNAL OF HIGH RESOLUTION CHROMATOGRAPHY & CHROMATOGRAPHY COMMUNICATIONS, 1988, 11 (08): : 554 - 559
  • [3] DETERMINATION OF 9 BETA-BLOCKERS IN SERUM BY MICELLAR ELECTROKINETIC CAPILLARY CHROMATOGRAPHY
    LUKKARI, P
    NYMAN, T
    RIEKKOLA, ML
    JOURNAL OF CHROMATOGRAPHY A, 1994, 674 (1-2) : 241 - 246
  • [4] DETERMINATION OF 10 BETA-BLOCKERS IN URINE BY MICELLAR ELECTROKINETIC CAPILLARY CHROMATOGRAPHY
    LUKKARI, P
    SIREN, H
    PANTSAR, M
    RIEKKOLA, ML
    JOURNAL OF CHROMATOGRAPHY, 1993, 632 (1-2): : 143 - 148
  • [5] Roles of organic modifiers in micellar electrokinetic capillary chromatography
    Liu, Z
    Zou, HF
    Zhang, YK
    HRC-JOURNAL OF HIGH RESOLUTION CHROMATOGRAPHY, 1998, 21 (04): : 234 - 240
  • [6] Roles of organic modifiers in micellar electrokinetic capillary chromatography
    Liu, Zhen
    Zou, Hanfa
    Zhang, Yukui
    Journal of High Resolution Chromatography, 1998, 21 (04): : 234 - 240
  • [7] Effects of organic modifiers on solute retention and electrokinetic migrations in micellar electrokinetic capillary chromatography
    Liu, Z
    Zou, HF
    Ye, ML
    Ni, JY
    Zhang, YK
    ELECTROPHORESIS, 1999, 20 (14) : 2898 - 2908
  • [8] INTRODUCTION OF MIGRATION INDEXES FOR IDENTIFICATION - CHIRAL SEPARATION OF SOME BETA-BLOCKERS BY USING CYCLODEXTRINS IN MICELLAR ELECTROKINETIC CAPILLARY CHROMATOGRAPHY
    SIREN, H
    JUMPPANEN, JH
    MANNINEN, K
    RIEKKOLA, ML
    ELECTROPHORESIS, 1994, 15 (06) : 779 - 784
  • [9] SEPARATION OF LIPOPHILIC COMPOUNDS BY MICELLAR ELECTROKINETIC CHROMATOGRAPHY WITH ORGANIC MODIFIERS
    OTSUKA, K
    HIGASHIMORI, M
    KOIKE, R
    KARUHAKA, K
    OKADA, Y
    TERABE, S
    ELECTROPHORESIS, 1994, 15 (10) : 1280 - 1283