DETERMINATION OF TOTAL PHOSPHORUS IN WATERS AND WASTEWATERS BY ONLINE MICROWAVE-INDUCED DIGESTION AND FLOW-INJECTION ANALYSIS

被引:45
|
作者
BENSON, RL
MCKELVIE, ID
HART, BT
HAMILTON, IC
机构
[1] MONASH UNIV,DEPT CHEM,CAULFIELD,VIC 3145,AUSTRALIA
[2] BHP RES,NEWCASTLE LABS,WALLSEND,NSW 2287,AUSTRALIA
关键词
FLOW INJECTION; SPECTROPHOTOMETRY; DIGESTION; FILTRATION; PHOSPHORUS; WATERS;
D O I
10.1016/0003-2670(94)80018-9
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
An on-line flow-injection system for the determination of total phosphorus in waters and wastewaters is described. Digestion was performed using a flow-through reactor in a modified domestic microwave oven, and the orthophosphate formed was detected as phosphomolybdenum blue after on-line filtration. The modifications and operation of the microwave complied with Australian Standards. Reagent and digestion conditions required to give optimal oxidation and hydrolysis of organic and condensed model species are reported. The flow-injection method gave complete recovery for almost all the model P compounds tested, with the exception of condensed phosphates. When the proposed method was applied to a series of wastewater samples, the results obtained were in close agreement with those from a conventional batch digestion method. The technique was rapid (7 samples h(-1), 4 replicates), with a linear range of interest for wastewaters 0-18 mg P 1(-1), and a detection limit of 0.09 mg P 1(-1).