FACTORS AFFECTING UNIFORMITY IN INTERPRETATION OF PLANAR TL-201 IMAGING IN A MULTICENTER TRIAL

被引:46
|
作者
WACKERS, FJT
BODENHEIMER, M
FLEISS, JL
BROWN, M
机构
[1] YALE UNIV,SCH MED,DEPT INTERNAL MED,CARDIOVASC MED SECT,NEW HAVEN,CT 06510
[2] LONG ISL JEWISH MED CTR,DIV ADULT CARDIOL,NEW HYDE PK,NY
[3] COLUMBIA UNIV,SCH PUBL HLTH,DIV BIOSTAT,NEW YORK,NY 10027
[4] UNIV ROCHESTER,HEART RES FOLLOW UP PROGRAM,ROCHESTER,NY 14627
关键词
D O I
10.1016/0735-1097(93)90226-Q
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Objectives. This study was designed te, assess factors affecting interobserver agreement in interpretation of planar thallium-201 stress imaging in the Multicenter Study on Silent Myocardial Ischemia (MSSMI). Background. Five hundred fifty-six planar thallium-201 images were interpreted in 24 clinical centers and in a Radionuclide Core Laboratory. The trial's Coordinating and Data Center observed that the participating clinical centers interpreted a significantly greater number of thallium-201 stress studies as abnormal (i.e., myocardial ischemia or scar) than the Core Laboratory, and overall agreement was poor (kappa 0.27). Methods. Agreement in image interpretation between clinical centers and the Radionuclide Core Laboratory was analyzed by kappa statistics. The reproducibility of the Core Laboratory results on 41 randomly selected test studies was excellent (kappa 0.77). In contrast, the reproducibility of interpretation in the clinical centers on their own studies was at best fair (kappa 0.45). It was hypothesized that the poor agreement and reproducibility in the clinical centers were caused by lack of standardization of image display and lack of objective criteria for image interpretation. To test the effect of standardization, 13 clinical investigators interpreted the same 41 test studies using 1) uniform image display, and 2) uniform quantification of images. Results. The agreement in interpretation between clinical investigators and the Radionuclide Core Laboratory improved modestly with uniformity of image display (kappa 0.57) but improved markedly (kappa 0.66) with quantitative circumferential profile analysis. Conclusions. Lack of standardization in image display and lack of objective criteria for interpretation of thallium-201 images are responsible for suboptimal reproducibility and poor interlaboratory agreement in the interpretation of thallium-201 stress imaging. The adoption of a uniformly accepted method for computer quantification of myocardial perfusion images is crucial to improve agreement in interpretation.
引用
收藏
页码:1064 / 1074
页数:11
相关论文
共 50 条
  • [1] A LOOK AT 15 YEARS OF PLANAR TL-201 IMAGING
    KAUL, S
    AMERICAN HEART JOURNAL, 1989, 118 (03) : 581 - 601
  • [2] A MULTICENTER TRIAL ON INTEROBSERVER AND INTRAOBSERVER REPRODUCIBILITY OF SEGMENTAL SCORING OF TL-201 PLANAR MYOCARDIAL IMAGING BEFORE AND AFTER REINJECTION
    BRAMBILLA, M
    INGLESE, E
    CANNIZZARO, G
    DONDI, M
    SARA, R
    ARRIGO, F
    TAROLO, GL
    JOURNAL OF NUCLEAR MEDICINE, 1994, 35 (04) : 601 - 608
  • [3] QUANTITATIVE PLANAR AND TOMOGRAPHIC TL-201 MYOCARDIAL PERFUSION IMAGING
    GARCIA, EV
    DEPUEY, EG
    DEPASQUALE, EE
    CARDIOVASCULAR AND INTERVENTIONAL RADIOLOGY, 1987, 10 (06) : 374 - 383
  • [4] QUANTITATIVE REPRODUCIBILITY OF REPEATED PLANAR TL-201 STRESS IMAGING
    WACKERS, FJT
    PIERI, PL
    MCMAHON, M
    DELLAVITTORIA, G
    TISSELLI, A
    MOSCATELLI, G
    SPINELLI, A
    RIVA, P
    MCKUSICK, KA
    STRAUSS, HW
    JOURNAL OF NUCLEAR MEDICINE, 1993, 34 (05) : P44 - P44
  • [5] MYOCARDIAL IMAGING WITH TL-201 AT REST AND EXERCISE - MULTICENTER STUDY
    RITCHIE, J
    ZARET, B
    STRAUSS, W
    PITT, B
    BERMAN, D
    SCHELBERT, H
    ASHBURN, W
    BERGER, H
    HAMILTON, G
    CIRCULATION, 1977, 56 (04) : 230 - 230
  • [6] TL-201 IMAGING OF TUMORS
    MCKILLOP, JH
    SEHWEIL, AM
    OMAR, YT
    ABDELDAYEM, HM
    AUSTRALIAN AND NEW ZEALAND JOURNAL OF MEDICINE, 1988, 18 (03): : 508 - 508
  • [7] TL-201 MYOCARDIAL IMAGING
    COWLEY, MJ
    MEDICAL COLLEGE OF VIRGINIA QUARTERLY, 1979, 15 (01): : 8 - 15
  • [8] TL-201 FOR MYOCARDIAL IMAGING
    POHOST, GM
    NUKLEARMEDIZIN, 1978, 17 (04) : 149 - 153
  • [9] THE VALUE OF TL-201 IMAGING
    POHOST, GM
    HENZLOVA, MJ
    NEW ENGLAND JOURNAL OF MEDICINE, 1990, 323 (03): : 190 - 192
  • [10] IMAGING OF THE MYOCARDIUM WITH TL-201
    HUMAN, GP
    DORMEHL, I
    SOUTH AFRICAN MEDICAL JOURNAL, 1981, 59 (15): : 524 - 528