CONSTRUCTION AND EVALUATION OF TUBULAR POTENTIOMETRIC DETECTORS SENSITIVE TO CHLORIDE, BROMIDE, AND IODIDE AND BASED ON HOMOGENEOUS CRYSTALLINE MEMBRANES

被引:29
|
作者
FERREIRA, IMPLVO [1 ]
LIMA, JLFC [1 ]
ROCHA, LSM [1 ]
机构
[1] FAC FARM PORTO,DEPT QUIM FIS,RUA ANIBAL CUNHA 164,P-4000 OPORTO,PORTUGAL
来源
关键词
D O I
10.1007/BF00323813
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The construction of tubular potentiometric detectors is described, without an inner reference solution, sensitive to chloride, bromide, and iodide. They are based on homogeneous crystalline membranes obtained by high-pressure pressing of AgX/Ag2S powdered mixtures (X = Cl, Br and 1). The respective operating characteristics are evaluated by comparison with conventionally-shaped electrodes with the same sensor. In general, the detectors showed similar operating characteristics as those of the corresponding conventional electrodes, namely the concentration range, in which they presented a Nernstian response, an excellent reproducibility, and an extremely rapid response rate which in some cases enabled the FIA manifolds to attain sampling rates of 360 samples/h.
引用
收藏
页码:314 / 319
页数:6
相关论文
共 8 条