A SYSTEM FOR NONINVASIVELY MEASURING BLOOD-PRESSURE ON A TREADMILL

被引:8
|
作者
AU, TCP
WEBSTER, JG
TAN, BT
机构
[1] Department of Electrical and Computer Engineering, University of Wisconsin, Madison, WI 53706-1691
关键词
BLOOD PRESSURE; KOROTKOFF SOUNDS; MOTION ARTIFACTS; SOUND SPECTRUM; TREADMILL EXERCISE; AUSCULTATORY METHOD; STRESS TESTING; SIGNAL TO NOISE RATIO; ELECTRONIC FILTER; FREQUENCY ANALYSIS;
D O I
10.1088/0967-3334/16/4/007
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
Measurement of signal to artifact ratio yielded optimal frequencies for a Korotkoff-based automatic system for measuring blood pressure on a treadmill. Maximal Korotkoff sounds occurred just above the crease of the elbow over the brachial artery. Output of a piezoelectric microphone and charge amplifier was measured during treadmill exercise for frequency bands from 8 to 57 Hz. An automatic system filtered output from 40 to 45 Hz, rectified it, then compared it to a threshold that had fixed and exponentially decaying components. For five subjects, the system decisions of systolic and diastolic pressures compared well with those of two observers using stethoscopes. The system shows promise for improved measurements of blood pressure during treadmill exercise.
引用
收藏
页码:273 / 285
页数:13
相关论文
共 50 条
  • [1] Related Work in Noninvasively Measuring Blood Pressure
    不详
    IEEE INTELLIGENT SYSTEMS, 2015, 30 (06) : 39 - 39
  • [2] MEASURING BLOOD-PRESSURE
    SWALES, JD
    BRITISH MEDICAL JOURNAL, 1993, 307 (6901): : 424 - 424
  • [3] MEASURING BLOOD-PRESSURE
    不详
    LANCET, 1976, 2 (7982): : 409 - 409
  • [4] MEASURING BLOOD-PRESSURE
    HARRIS, A
    BRITISH MEDICAL JOURNAL, 1975, 4 (5999): : 759 - 759
  • [5] MEASURING BLOOD-PRESSURE
    不详
    LANCET, 1966, 1 (7434): : 414 - +
  • [6] MEASURING BLOOD-PRESSURE
    不详
    BRITISH MEDICAL JOURNAL, 1975, 4 (5993): : 366 - 367
  • [7] ACCURACY OF NONINVASIVELY MEASURED BLOOD-PRESSURE FROM A FINGER
    DAS, G
    ELADASARI, B
    CZERNEK, P
    CLINICAL RESEARCH, 1993, 41 (01): : A12 - A12
  • [8] A system for non-invasively measuring blood pressure on a treadmill
    Dept. Electrical/Computer Engineer., University of Wisconsin, 1415 Engineering Drive, Madison, WI 53706-1691, United States
    PHYSIOL. MEAS., 4 (273-285):
  • [9] ERRORS IN ASSESSMENT OF BLOOD-PRESSURE - BLOOD-PRESSURE MEASURING TECHNIQUE
    CAMPBELL, NRC
    MCKAY, DW
    CHOCKALINGAM, A
    FODOR, JG
    CANADIAN JOURNAL OF PUBLIC HEALTH-REVUE CANADIENNE DE SANTE PUBLIQUE, 1994, 85 : S18 - S21
  • [10] INDIRECT BLOOD-PRESSURE MEASURING SYSTEM WITH DIGITAL READOUT
    EATON, JCR
    HENSON, TA
    BRITISH JOURNAL OF PHARMACOLOGY, 1975, 54 (02) : P282 - P282