LIMITATIONS OF ULTRASONOGRAPHY IN SURVEILLANCE OF SMALL ABDOMINAL AORTIC-ANEURYSMS

被引:112
|
作者
ELLIS, M [1 ]
POWELL, JT [1 ]
GREENHALGH, RM [1 ]
机构
[1] CHARING CROSS & WESTMINSTER MED SCH,DEPT SURG,FULHAM PALACE RD,LONDON W6 8RF,ENGLAND
关键词
D O I
10.1002/bjs.1800780529
中图分类号
R61 [外科手术学];
学科分类号
摘要
The repeatability, observer bias and instrument bias of aortic diameter measurement by ultrasonography, were investigated in ten patients with small (3-6 cm by computed tomography) infrarenal abdominal aortic aneurysm. The repeatability of maximum aortic diameter measurement by ultrasonography was much better for anterior-posterior than transverse diameter, with coefficients of repeatability 3.0-7.5 mm and 10-15 mm respectively. The repeatability of suprarenal aortic diameter measurement was poor. Surprisingly, maximum diameter using ultrasonography was larger than that using computed tomography, the difference being least for anterior-posterior measurements. At best a single, experienced observer, using the same instrument may provide aortic diameters using ultrasonography accurate to within 5 mm, but more commonly such aortic diameter is only accurate to within 8 mm.