ECHO-PLANAR IMAGING;
SHIMMING;
OVERSAMPLING;
POINT SPREAD FUNCTION;
FUNCTIONAL IMAGING;
D O I:
10.1016/0730-725X(94)92462-7
中图分类号:
R8 [特种医学];
R445 [影像诊断学];
学科分类号:
1002 ;
100207 ;
1009 ;
摘要:
The high speed magnetic resonance imaging technique of Echo-Planar imaging (EPI) has been developing since 1977 using specialized imaging systems. Recent experiments in functional magnetic resonance imaging (FMRI) using EPI on specialized MR systems have generated increased interest in the EPI sequence. Presented here is an implementation of the EPI sequence on a large bore, human scale spectrometer using widely available standard hardware. Existing EPI methods and concepts were amalgamated to obtain images without hardware modifications, Spatial distortion of Echo-Planar images by internal field gradients within the brain were minimized by active shimming on each subject. Nonlinear sampling artifacts caused by nonideal gradient switching were addressed by oversampling the signal and simple extraction of data points corresponding to the ideal sampling pattern. Multislice Echo-Planar images with functional contrast have been obtained at 6 x 3 mm resolution using these methods.
机构:
Massachusetts Gen Hosp, Dept Radiol, AA Martinos Ctr, Charlestown, MA USA
MIT, Harvard Mit Div Hlth Sci & Technol, Cambridge, MA 02139 USAUniv Duisburg Essen, Erwin L Hahn Inst Magnet Resonance Imaging, D-45141 Essen, Germany
Witzel, T.
Wald, L. L.
论文数: 0引用数: 0
h-index: 0
机构:
Massachusetts Gen Hosp, Dept Radiol, AA Martinos Ctr, Charlestown, MA USA
MIT, Harvard Mit Div Hlth Sci & Technol, Cambridge, MA 02139 USAUniv Duisburg Essen, Erwin L Hahn Inst Magnet Resonance Imaging, D-45141 Essen, Germany