ELECTROCHEMICAL MEASUREMENT OF CARBON-MONOXIDE IN BREATH - INTERFERENCE BY HYDROGEN

被引:16
|
作者
VREMAN, HJ
MAHONEY, JJ
STEVENSON, DK
机构
[1] Department of Pediatrics and Pulmonary Physiology, Stanford University School of Medicine, Stanford
来源
基金
美国国家卫生研究院;
关键词
GAS CHROMATOGRAPHY; CO-OXIMETRY; CARBOXYHEMOGLOBIN; ELECTROCHEMICAL ANALYSIS; CO;
D O I
10.1016/0960-1686(93)90049-5
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
The purpose of this study was to determine the concentration of carbon monoxide (CO) in blood (COHb) and breath to demonstrate that breath hydrogen (H2) can be a significant interferant. For this purpose, we measured blood COHb with CO-oximetry and breath CO with an electrochemical analyzer. In addition, the samples were analyzed by gas chromatography (GC). The concentration of CO in breath, collected with a Priestley tube after a 20 s breath hold, from healthy, nonsmoking adult males (n = 20) and females (n = 10) had a mean +/- SD (range) of 2.6 +/- 0.4 ppm (2.0-3.9), respectively, when measured by GC. However, these same samples when measured with an electrochemical (EC) analyzer showed elevated CO values of 4.7 +/- 2.9 ppm (2.6-17.6). The concentration of H-2, a prominent trace gas in breath known to interfere with EC analyzers, correlated strongly with the observed EC analyzer response [EC (ppm CO) = 0.336 H-2 (PPM) + 1.93, r2 = 0.98]. The EC analyzer was linear for H-2 concentrations up to 40 ppm, with a sensitivity of 0.035 V ppm-1.The analyzer sensitivity to CO was 0.10 V ppm-1. Blood from this population showed COHb concentrations of 0.56 +/- 0.11% (0.40-0.97), as measured by GC, but elevated values were found when measured by CO-oximeter (Ciba Coming Diagnostics Corp., Models 2500 and 270), 1.3 +/- 0.2% (1.1-1.6) and 1.0 +/- 0.3% (0.1-1.6), respectively. When breath CO was compared to blood COHb, only measurements by GC significantly correlated [COHb% = 0.241 CO(ppm)-0.076, r2 = 0.78]. We conclude that, relative to quantitative analysis by GC, (1) EC analyzers are susceptible to H-2 interference that cause falsely elevated CO measurements, and (2) CO-oximeters overestimate COHb concentrations in the range typical for healthy nonsmokers.
引用
收藏
页码:2193 / 2198
页数:6
相关论文
共 50 条
  • [1] ELECTROCHEMICAL TECHNIQUE FOR MEASUREMENT OF CARBON-MONOXIDE
    BAY, HW
    BLURTON, KF
    SEDLAK, JM
    VALENTINE, AM
    [J]. ANALYTICAL CHEMISTRY, 1974, 46 (12) : 1837 - 1839
  • [2] SEMIPORTABLE ELECTROCHEMICAL INSTRUMENT FOR DETERMINING CARBON-MONOXIDE IN BREATH
    VREMAN, HJ
    STEVENSON, DK
    OH, W
    FANAROFF, AA
    WRIGHT, LL
    LEMONS, JA
    WRIGHT, E
    SHANKARAN, S
    TYSON, JE
    KORONES, SB
    BAUER, CR
    STOLL, BJ
    PAPILE, LA
    DONOVAN, EF
    EHRENKRANZ, RA
    [J]. CLINICAL CHEMISTRY, 1994, 40 (10) : 1927 - 1933
  • [3] ACUTE EFFECT OF SMOKING ON REBREATHING CARBON-MONOXIDE, BREATH-HOLD CARBON-MONOXIDE AND ALVEOLAR OXYGEN
    KIRKHAM, AJT
    GUYATT, AR
    CUMMING, G
    [J]. CLINICAL SCIENCE, 1988, 75 (04) : 371 - 373
  • [4] A PLANAR ELECTROCHEMICAL CARBON-MONOXIDE SENSOR
    YASUDA, A
    YAMAGA, N
    DOI, K
    FUJIOKA, T
    KUSANAGI, S
    [J]. JOURNAL OF THE ELECTROCHEMICAL SOCIETY, 1992, 139 (04) : 1091 - 1095
  • [5] CARBON-MONOXIDE IN AIR AND BREATH OF EMPLOYEES IN AN UNDERGROUND OFFICE
    WALLACE, LA
    [J]. JOURNAL OF THE AIR POLLUTION CONTROL ASSOCIATION, 1983, 33 (07): : 678 - 682
  • [6] CARBON-MONOXIDE IN BREATH IN RELATION TO SMOKING AND CARBOXYHEMOGLOBIN LEVELS
    WALD, NJ
    IDLE, M
    BOREHAM, J
    BAILEY, A
    [J]. THORAX, 1981, 36 (05) : 366 - 369
  • [7] A BREATH OF CARBON-MONOXIDE REDUCES CHEMORECEPTOR DISCHARGE IN THE CAT
    NYE, PCG
    TORRANCE, RW
    [J]. JOURNAL OF PHYSIOLOGY-LONDON, 1980, 307 (OCT): : P46 - P47
  • [8] ADSORPTION OF CARBON-MONOXIDE AND HYDROGEN ON GRAPHITE
    BEITEL, GA
    [J]. JOURNAL OF VACUUM SCIENCE & TECHNOLOGY, 1972, 9 (01): : 370 - &
  • [9] ADSORPTION OF HYDROGEN AND CARBON-MONOXIDE ON NICKEL
    CRAIG, JH
    [J]. SURFACE SCIENCE, 1984, 141 (01) : L291 - L294
  • [10] ADSORPTION OF HYDROGEN AND CARBON-MONOXIDE ON PLATINUM
    CRAIG, JH
    [J]. SURFACE SCIENCE, 1981, 111 (02) : L695 - L700