EQUIPMENT FAILURE - AN ANALYSIS OF 2000 INCIDENT REPORTS
被引:39
|作者:
WEBB, RK
RUSSELL, WJ
KLEPPER, I
RUNCIMAN, WB
机构:
[1] UNIV ADELAIDE,DEPT ANAESTHESIA & INTENS CARE,ADELAIDE,SA 5001,AUSTRALIA
[2] ROYAL ADELAIDE HOSP,ADELAIDE,SA 5000,AUSTRALIA
关键词:
ANESTHESIA;
EQUIPMENT;
INFUSION PUMP;
MONITOR;
VAPORIZER;
REGULATOR;
HUMIDIFIER;
BLOOD WARMER;
TOURNIQUET;
GAS SUPPLY;
ELECTRICITY SUPPLY;
D O I:
10.1177/0310057X9302100533
中图分类号:
R614 [麻醉学];
学科分类号:
100217 ;
摘要:
Of the first 2000 incidents reported to the Australian Incident Monitoring Study, 177 (9%) were due to ''pure'' equipment failure according to pre-defined criteria. Of these 107 (60%) involved anaesthetic equipment, 42 (24%) involved monitors, 17 (10%) other theatre equipment and 11 (6%) the gas or electricity supply. Ninety-seven (55% of the 177) were potentially life-threatening; of these two-thirds would be detected by the array of monitors recommended by the Australian and New Zealand College of Anaesthetists and all but 9 of the remainder would be handled by application of the crisis management algorithm recommended elsewhere in this symposium. Of the 9 remaining, 2 were electrical shock, 3 overheating of a humidifier or blood warmer, 2 the unavailability of a spare laryngoscope and 1 the consequence of a power failure. Meticulous adherence to the equipment checking and monitoring guidelines of the Australian and New Zealand College of Anaesthetists and application of a suitable crisis management algorithm should protect the patient from potentially life-threatening equipment failure in virtually all cases except electric shock, power failure and overheating of warming devices.
引用
收藏
页码:673 / 677
页数:5
相关论文