AN ASSESSMENT OF THE CEREBROTRAC-2500 FOR CONTINUOUS MONITORING OF CEREBRAL FUNCTION IN THE INTENSIVE-CARE UNIT

被引:17
|
作者
SHEARER, ES
OSULLIVAN, EP
HUNTER, JM
机构
[1] University Department of Anaesthesia, Royal Liverpool Hospital, Liverpool, L69 3BX, Prescot Street
关键词
EQUIPMENT; CEREBROTRAC-2500; MONITORING; CEREBRAL FUNCTION;
D O I
10.1111/j.1365-2044.1991.tb09772.x
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Experience of the use of the Cerebrotrac 2500 EEG monitor in 17 patients subjected to artificial ventilation in an intensive care unit is reported; seven were receiving continuous sedation with morphine, midazolam and propofol singly or in combination and 10 received both sedation and the neuromuscular blocking agent, atracurium. The processed EEG patterns could not be precisely correlated with a standard clinical scoring system but were useful in determining the adequacy of sedation, particularly when a muscle relaxant was used. The monitor also shows considerable promise in the management of the paralysed patient with widespread convulsive activity in whom ischaemic brain damage may be occurring from epileptiform activity in the absence of any clinical manifestation. The ability to detect cerebral irritability or isolated epileptiform discharges using this apparatus is, however, questionable. The equipment was easy to use and robust; the running costs were 9.5p per hour. Copyright © 1991, Wiley Blackwell. All rights reserved
引用
收藏
页码:750 / 755
页数:6
相关论文
共 50 条
  • [1] CONTINUOUS ELECTROPHYSIOLOGIC MONITORING OF CEREBRAL FUNCTION IN THE PEDIATRIC INTENSIVE-CARE UNIT
    TALWAR, D
    TORRES, F
    [J]. PEDIATRIC NEUROLOGY, 1988, 4 (03) : 137 - 147
  • [2] EVALUATION OF THE CEREBRO-TRAC-2500 FOR MONITORING OF CEREBRAL FUNCTION IN THE NEONATAL INTENSIVE-CARE
    EATON, DM
    TOET, M
    LIVINGSTON, J
    SMITH, I
    LEVENE, M
    [J]. NEUROPEDIATRICS, 1994, 25 (03) : 122 - 128
  • [3] MONITORING OF CEREBRAL FUNCTION IN SEDATED AND VENTILATED PATIENTS IN AN INTENSIVE-CARE UNIT
    LEHMKUHL, P
    LIPS, U
    PICHLMAYR, I
    [J]. ELECTROENCEPHALOGRAPHY AND CLINICAL NEUROPHYSIOLOGY, 1986, 63 (01): : P22 - P22
  • [4] THE CLINICAL-VALUE OF MONITORING CEREBRAL FUNCTION IN AN INTENSIVE-CARE UNIT
    BOZZAMARRUBINI, M
    BEFFAGNA, B
    LIVINI, M
    MOMBELLI, F
    PROCACCIO, F
    SOTTILI, S
    [J]. CRITICAL CARE MEDICINE, 1981, 9 (03) : 241 - 241
  • [5] MONITORING NEUROMUSCULAR FUNCTION IN THE INTENSIVE-CARE UNIT
    VIBYMOGENSEN, J
    [J]. INTENSIVE CARE MEDICINE, 1993, 19 : S74 - S79
  • [6] CEREBRAL FUNCTION MONITORING IN NEONATAL INTENSIVE-CARE
    DODD, KL
    [J]. ARCHIVES OF DISEASE IN CHILDHOOD, 1978, 53 (08) : 687 - 687
  • [7] CONTINUOUS ECG MONITORING IN A SURGICAL INTENSIVE-CARE UNIT
    LEWIS, FJ
    QUINN, ML
    [J]. CRITICAL CARE MEDICINE, 1977, 5 (02) : 73 - 75
  • [8] ASSESSMENT OF DIAPHRAGMATIC FUNCTION IN THE INTENSIVE-CARE UNIT
    DUREUIL, B
    AUBIER, M
    [J]. INTENSIVE CARE MEDICINE, 1988, 14 (02) : 83 - 85
  • [9] THE CEREBRAL FUNCTION MONITOR IN A TRAUMATIC INTENSIVE-CARE UNIT
    ROMANO, P
    MARGENET, A
    ROUJAS, F
    HUGUENARD, P
    [J]. CRITICAL CARE MEDICINE, 1981, 9 (03) : 240 - 240
  • [10] MONITORING IN THE INTENSIVE-CARE UNIT
    ADAMS, AP
    [J]. ANAESTHESIST, 1985, 34 (06): : 319 - 319