Effect of variable circuit flow rate during the expiratory phase on CO2 elimination

被引:0
|
作者
Keszler, Peter A. [1 ,2 ]
Nagaraj, Pankaj [1 ,2 ]
Abubakar, Kabir [1 ,2 ]
Keszler, Martin [3 ]
机构
[1] Georgetown Univ, Washington, DC USA
[2] Georgetown Univ Hosp, Washington, DC 20007 USA
[3] Brown Univ, Women & Infants Hosp Rhode Isl, Providence, RI 02912 USA
关键词
mechanical ventilation; newborn; bench study; re-breathing; expiratory flow rate; end-tidal CO2;
D O I
10.2147/RRN.S28346
中图分类号
R72 [儿科学];
学科分类号
100202 ;
摘要
Background: Some continuous flow infant ventilators allow independent setting of inspiratory and expiratory circuit flow rate. In the Drager Babylog 8000+ ventilator, this is called "variable inspiratory, variable expiratory flow" (VIVE). Some clinicians believe that lower expiratory flow decreases expiratory resistance. The minimum expiratory flow rate needed to avoid -re-breathing of carbon dioxide (CO2) has never been established. Objective: We sought to determine if re-breathing becomes evident at the lowest possible expiratory flow rate setting of 1 L/min. Design/methods: We conducted a bench study using end-tidal CO2 (ETCO2) measurement and a 45 mL (90 mL for the "term" model) test lung pre-filled with 100% CO2. We previously showed that the time needed for ETCO2 to be eliminated from the lung is a highly reproducible indicator of efficiency of ventilation. Re-breathing would thus be identified by an increase in the time required for the CO2 to be washed out from the test lung at stable settings of rate and tidal volume (V-T). Using a Babylog 8000+ ventilator in volume guarantee mode with VIVE and a standard ventilator circuit, we tested the effect of decreasing expiratory flow rate under conditions simulating three sizes of patients: extremely low birth weight infant, wt = 600 g (V-T = 3.5 mL, respiratory rate (RR) = 60 breaths min(-1), minute ventilation (MV) = 210 mL/min, expiratory flow rate = 3 L/min, 2 L/min, and 1 L/min), very low birth weight infant, wt = 1.5 kg (V-T = 7 mL, RR = 60 breaths min(-1), MV = 420 mL/min, expiratory flow rate = 4 L/min, 3 L/min, 2 L/min, and 1 L/min), and term infant, wt = 3.6 kg (V-T = 16 mL, RR = 60 breaths min(-1), MV = 960 mL/min, expiratory flow rate = 5 L/min, 4 L/min, 3 L/min, 2 L/min, and 1 L/min). Each measurement was repeated four times and the mean values were compared by analysis of variance for repeated measures. Results: The CO2 elimination times were always within 10% of each other with each repetition. There was no significant increase in the elimination time even at the lowest flows. Conclusion: Clinically significant re-breathing does not occur even with expiratory flow rate at the lowest possible setting of 1 L/min. VIVE is safe to use, although its clinical utility has not been established.
引用
收藏
页码:33 / 37
页数:5
相关论文
共 50 条
  • [1] MAXIMUM EXPIRATORY FLOW AND ESTIMATED CO2 ELIMINATION IN LIQUID-VENTILATED DOGS LUNGS
    SCHOENFISCH, WH
    KYLSTRA, JA
    [J]. JOURNAL OF APPLIED PHYSIOLOGY, 1973, 35 (01) : 117 - 121
  • [2] FLUOROCARBON VENTILATION - MAXIMAL EXPIRATORY FLOWS AND CO2 ELIMINATION
    KOEN, PA
    WOLFSON, MR
    SHAFFER, TH
    [J]. PEDIATRIC RESEARCH, 1988, 24 (03) : 291 - 296
  • [3] The effect of light level, CO2 flow rate, and anesthesia on the stress response of mice during CO2 euthanasia
    Karin Powell
    Kelly Ethun
    Douglas K. Taylor
    [J]. Lab Animal, 2016, 45 : 386 - 395
  • [4] The effect of light level, CO2 flow rate, and anesthesia on the stress response of mice during CO2 euthanasia
    Powell, Karin
    Ethun, Kelly
    Taylor, Douglas K.
    [J]. LAB ANIMAL, 2016, 45 (10) : 386 - 395
  • [5] INFLUENCE OF EXPIRATORY FLOW PATTERN ON CO2 EXCHANGE
    FEENSTRA, BWA
    LUIJENDIJK, SCM
    JANSEN, JRC
    VANROOYEN, W
    BRUINING, HA
    [J]. BULLETIN EUROPEEN DE PHYSIOPATHOLOGIE RESPIRATOIRE-CLINICAL RESPIRATORY PHYSIOLOGY, 1985, 21 : 9 - 9
  • [6] RATE OF ELIMINATION OF EXCESS CO2 IN HUMANS
    CHONAN, T
    ELHEFNAWY, AM
    SIMONETTI, OP
    CHERNIACK, NS
    [J]. RESPIRATION PHYSIOLOGY, 1988, 73 (03): : 379 - 394
  • [7] EFFECT OF BREATHING PATTERNS ON CO2 ELIMINATION DURING LIQUID VENTILATION
    KOEN, PA
    SHAFFER, TH
    [J]. FEDERATION PROCEEDINGS, 1978, 37 (03) : 868 - 868
  • [8] Effect of the Flow Rate on the Relative Permeability Curve in the CO2 and Brine System for CO2 Sequestration
    Jeong, Gu Sun
    Ki, Seil
    Lee, Dae Sung
    Jang, Ilsik
    [J]. SUSTAINABILITY, 2021, 13 (03) : 1 - 14
  • [9] PEEP valves during manual hyperinflation - Effect on expiratory flow rate
    Chan, P
    Paratz, J
    [J]. INTENSIVE CARE MEDICINE, 2002, 28 : S196 - S196
  • [10] Transient flow in wellbores and phase transition of CO2 during formation supercritical CO2 invasion
    Dou, Liangbin
    Zhang, Ming
    Bi, Gang
    Li, Tiantai
    [J]. ENERGY SCIENCE & ENGINEERING, 2019, 7 (02): : 323 - 337