Haptic Breakage after Transscleral Fixation of a Single-Piece Acrylic Intraocular Lens

被引:4
|
作者
Tanaka, Hiroki [1 ,2 ]
Fujinami, Kaoru [1 ,2 ,3 ,4 ]
Watanabe, Ken [1 ,2 ]
Noda, Toru [1 ,2 ]
Akiyama, Kunihiko [1 ,2 ]
机构
[1] Natl Hosp Org, Natl Tokyo Med Ctr, Dept Ophthalmol, Tokyo, Japan
[2] Keio Univ, Natl Tokyo Med Ctr, Natl Inst Sensory Organs, Lab Visual Physiol, Tokyo, Japan
[3] Keio Univ, Sch Med, Dept Ophthalmol, Tokyo, Japan
[4] UCL Inst Ophthalmol, London, England
来源
CASE REPORTS IN OPHTHALMOLOGY | 2014年 / 5卷 / 02期
关键词
Intraocular lens; Acrylic intraocular lens; Intraocular lens damage; Transscleral fixation;
D O I
10.1159/000365350
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: To report the case of a patient with a damaged haptic of an acrylic intraocular lens (IOL) after transscleral IOL fixation. Methods: This is a retrospective and descriptive case report. Results: A 40-year-old man presented with sudden visual loss in the left eye. He had undergone phacoemulsification/aspiration and IOL implantation 3 years previously and IOL repositioning with transscleral fixation 2 years before the initial visit. His visual acuity was 0.3 in the left eye due to corneal edema caused by a foreign body, i.e., the severed haptic tip of the single-piece acrylic IOL, which was surgically removed. The IOL itself was tilted and therefore explanted. The surface of the tip of the haptic suggested that the monofilament suture thread had exerted continuous force across the haptic. Conclusions: The haptics of acrylic IOLs can be damaged after transscleral fixation. (C) 2014 S. Karger AG, Basel
引用
收藏
页码:212 / 216
页数:5
相关论文
共 50 条
  • [1] Breakage of the haptic of a single-piece acrylic intraocular lens after transscleral fixation
    Choi, Suk Kyue
    Kim, Jin Hyoung
    Lee, Doh
    Lee, Jong Hyun
    [J]. ACTA OPHTHALMOLOGICA, 2010, 88 (08) : e346 - e347
  • [2] Transscleral fixation of a single-piece hydrophilic foldable acrylic intraocular lens
    Taskapili, Muhittin
    Gulkilik, Gokhan
    Engin, Gunay
    Kocabora, Mehmet Selim
    Yilmazli, Cemil
    Ozsutcu, Mustafa
    Kucuksahin, Hasan
    [J]. CANADIAN JOURNAL OF OPHTHALMOLOGY-JOURNAL CANADIEN D OPHTALMOLOGIE, 2007, 42 (02): : 256 - 261
  • [3] Transscleral fixation of single-piece foldable acrylic lens with eyelets at the optic-haptic junction
    Yang, Jee Myung
    Yoon, Kyung-Chul
    Ji, Yong-Sok
    [J]. CANADIAN JOURNAL OF OPHTHALMOLOGY-JOURNAL CANADIEN D OPHTALMOLOGIE, 2015, 50 (05): : 367 - 372
  • [4] Transscleral Fixation of Single-Piece Hydrophilic Acrylic Lenses With No Eyelets
    Taskapili, Muhittin
    Gulkilik, Gokhan
    Kocabora, Mehmet Selim
    Nilay, Kandemir
    [J]. OPHTHALMIC SURGERY LASERS & IMAGING, 2009, 40 (04) : 434 - 436
  • [5] Single-piece acrylic intraocular lens in pediatric cataracts
    Edelson, C
    Dureau, P
    Adjadj, E
    Caputo, G
    de Meux, PD
    [J]. INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2004, 45 : U165 - U165
  • [6] Single-piece acrylic intraocular lens implantation in children
    Trivedi, RH
    Wilson, ME
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2003, 29 (09): : 1738 - 1743
  • [7] Single-piece acrylic intraocular lens implantation in children
    Sony, P
    Khokhar, S
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2004, 30 (09): : 1818 - 1819
  • [8] Haptic tuck for reverse optic capture of a single-piece acrylic toric or other single-piece acrylic intraocular lenses
    Gimbel, Howard V.
    Marzouk, Hala A.
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2019, 45 (02): : 125 - 129
  • [9] Scleral fixation of a single-piece multifocal intraocular lens
    Can, Ertugrul
    Basaran, M. Resat
    Gul, Adem
    [J]. EUROPEAN JOURNAL OF OPHTHALMOLOGY, 2013, 23 (02) : 249 - 251
  • [10] Migration of lens epithelial cells through haptic root of single-piece acrylic-foldable intraocular lens
    Sugita, M
    Kato, S
    Sugita, G
    Oshika, T
    [J]. AMERICAN JOURNAL OF OPHTHALMOLOGY, 2004, 137 (02) : 377 - 379