PRECOLUMN DEPROTEINIZATION METHOD OF HUMAN PLASMA USING A HYDROXYAPATITE CARTRIDGE FOR HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHIC AMINO-ACID-ANALYSIS

被引:1
|
作者
IWASE, H
ONO, I
机构
[1] Ajinomoto Co., Inc, Central Research Laboratories, Suzuki-cho
关键词
PRECOLUMN DEPROTEINIZATION METHOD; HYDROXYAPATITE CARTRIDGE; HIGH-PERFORMANCE LIQUID CHROMATOGRAPHY; NINHYDRIN DERIVATIZATION; AMINO ACID;
D O I
10.2116/analsci.11.73
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Determination of amino acids in human plasma was performed by postcolumn ninhydrin derivatization liquid chromatography using a hydroxyatite cartridge for precolumn deproteinization. This method was suitable for the determination of amino acids in human plasma. The proposed method was simple, rapid (deproteinization time: within 1 min) and reproducible (relative standard deviation: ca. 0.4 to 5.4% (n=3)). Average recovery of 22 amino acids was over 80%. It took about 2 h to elute the amino acids in human plasma on the chromatogram completely. Protein-binding tryptophan was also determined by the proposed method.
引用
收藏
页码:73 / 77
页数:5
相关论文
共 50 条