RAPID ONLINE PRECOLUMN HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHIC METHOD FOR THE DETERMINATION OF BENOMYL, CARBENDAZIM AND ALDICARB SPECIES IN DRINKING-WATER

被引:13
|
作者
MARVIN, CH
BRINDLE, ID
HALL, CD
CHIBA, M
机构
[1] AGR CANADA,RES STN,VINELAND STN L0R 2E0,ONTARIO,CANADA
[2] BROCK UNIV,DEPT CHEM,ST CATHARINES L2S 3A1,ONTARIO,CANADA
[3] ONTARIO MINIST ENVIRONM,REXDALE M9W 5L1,ONTARIO,CANADA
来源
JOURNAL OF CHROMATOGRAPHY | 1991年 / 555卷 / 1-2期
关键词
D O I
10.1016/S0021-9673(01)87174-0
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A reversed-phase high-performance liquid chromatographic (HPLC) method has been developed for the determination of trace concentrations of benomyl, carbendazim, aldicarb, aldicarb sulphoxide and aldicarb sulphone in drinking water. A 10-ml sample of water is passed through a 3-cm precolumn, packed with 5-mu-m C8 sorbent, at a flow-rate of 5 ml/min. The HPLC system is then switched to an acetonitrile-water gradient elution program. The preconcentrated analytes are eluted from, and separated by, the 3-cm C8 precolumn and determined by UV absorption. The total analytical time is 25 min. The lowest detectable concentrations are in the range of 2.5 . 10(-9) - 11.0 . 10(-9) g/ml for the five analytes investigated with 10 ml of sample.