Calibration of tri-axial MEMS accelerometers in the low-frequency range - Part 1: comparison among methods

被引:21
|
作者
D'Emilia, Giulio [1 ]
Gaspari, Antonella [1 ]
Mazzoleni, Fabrizio [2 ]
Natale, Emanuela [1 ]
Schiavi, Alessandro [2 ]
机构
[1] Univ Aquila, Dept Ind & Informat Engn & Econ, I-67100 Laquila, Italy
[2] Natl Inst Metrol Res, INRiM, I-10135 Turin, Italy
关键词
D O I
10.5194/jsss-7-245-2018
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Two alternative experimental procedures for the calibration of tri-axial accelerometers have been compared with traditional methods, performed according the procedures stated in the standard ISO 16063-21. Standard calibration is carried out by comparison with a laser Doppler vibrometer (LDV), used as a primary reference transducer. The main sensitivities have been investigated and, where applicable, also transverse ones. Many aspects have been evaluated: the hypotheses about transverse sensitivities, the simplicity of the procedure, the number of measurements needed, and the effect of typology of transducer, depending on electrical and geometrical contributions. Two different accelerometers have been tested, a piezo-electric accelerometer and a capacitive MEMS accelerometer. A low-frequency range of vibration has been investigated, 3 and 6 Hz, with amplitude of acceleration ranging from 2 to 20 ms(-2). A satisfactory reproducibility of methods has been verified, with percentage differences less than 2.5%. Anyway, pros and cons of each method are also discussed with reference to their possible use for easy and quick calibration of low-cost tri-axial accelerometers.
引用
收藏
页码:245 / 257
页数:13
相关论文
共 8 条
  • [1] Calibration of tri-axial MEMS accelerometers in the low-frequency range - Part 2: Uncertainty assessment
    D'Emilia, Giulio
    Gaspari, Antonella
    Mazzoleni, Fabrizio
    Natale, Emanuela
    Schiavi, Alessandro
    JOURNAL OF SENSORS AND SENSOR SYSTEMS, 2018, 7 (01) : 403 - 410
  • [2] Amplitude-phase calibration of tri-axial accelerometers in the low-frequency range by a LDV
    D'Emilia, Giulio
    Gaspari, Antonella
    Natale, Emanuela
    JOURNAL OF SENSORS AND SENSOR SYSTEMS, 2019, 8 (01) : 223 - 231
  • [3] A simple method for amplitude/phase calibration of tri-axial accelerometers in the low frequency range
    D'Emilia, G.
    Gaspari, A.
    Natale, E.
    13TH INTERNATIONAL CONFERENCE ON VIBRATION MEASUREMENTS BY LASER AND NONCONTACT TECHNIQUES, 2018, 2018, 1149
  • [4] A new type of tri-axial accelerometers with high dynamic range MEMS for earthquake early warning
    Peng, Chaoyong
    Chen, Yang
    Chen, Quansheng
    Yang, Jiansi
    Wang, Hongti
    Zhu, Xiaoyi
    Xu, Zhigiang
    Zheng, Yu
    COMPUTERS & GEOSCIENCES, 2017, 100 : 179 - 187
  • [5] A Novel Tri-Axial MEMS Gyroscope Calibration Method over a Full Temperature Range
    Yang, Haotian
    Zhou, Bin
    Wang, Lixin
    Xing, Haifeng
    Zhang, Rong
    SENSORS, 2018, 18 (09)
  • [6] Monocular vision-based dynamic calibration method for determining the sensitivities of low-frequency tri-axial vibration sensors
    Liu, Wenfeng
    Cai, Chenguang
    Liu, Zhihua
    Yang, Ming
    Zuo, Shengnan
    Wang, Peicheng
    OPTICS EXPRESS, 2024, 32 (06) : 10130 - 10145
  • [7] MODIFIED SENSOR ERROR MODEL FOR STATIC CALIBRATION OF A LOW-COST TRI-AXIAL MEMS ACCELEROMETER
    Uzair, Muhammad
    Khan, Ali F.
    Khurshid, Khawar
    Jeon, Byeungwoo
    INTERNATIONAL JOURNAL OF ROBOTICS & AUTOMATION, 2018, 33 (03): : 233 - 238
  • [8] Validity of using tri-axial accelerometers to measure human movement - Part II: Step counts at a wide range of gait velocities
    Fortune, Emma
    Lugade, Vipul
    Morrow, Melissa
    Kaufman, Kenton
    MEDICAL ENGINEERING & PHYSICS, 2014, 36 (06) : 659 - 669