DEUTERIUM-OXIDE AS A REAGENT FOR THE MODIFICATION OF MASS-SPECTRA IN ELECTROSPRAY MICROCOLUMN LIQUID-CHROMATOGRAPHY MASS-SPECTROMETRY

被引:25
|
作者
KARLSSON, KE
机构
来源
JOURNAL OF CHROMATOGRAPHY | 1993年 / 647卷 / 01期
关键词
D O I
10.1016/0021-9673(93)83321-I
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Microcolumn liquid chromatography was used in combination with a quadrupole mass spectrometer equipped with an electrospray ionization interface. Deuterium oxide was used as a reagent to induce peak shifts in mass spectra. The peak shift obtained gives the number of heteroatoms carrying hydrogen atoms that are present in a molecule. Examples relating to amino alcohols, peptides and sugars are given. Two different modes of operation were investigated; postcolumn addition of the reagent to the eluent from the column and the use of deuterium oxide as a mobile phase component. The latter technique makes use of a dynamic exchange reaction on-column. Exchanged hydrogen atoms are carried away from the migrating zone (as H2HO or H2O) containing the target compound, leading to a high yield of the deuterated compound. On-column exchange was found to be more useful, especially if the compounds contained several exchangeable hydrogen atoms. No significant changes were observed with regard to chromatographic selectivity or efficiency
引用
收藏
页码:31 / 38
页数:8
相关论文
共 50 条