Accuracy of automated blood pressure measurements in the presence of atrial fibrillation: systematic review and meta-analysis

被引:0
|
作者
Christopher E. Clark
Sinead T. J. McDonagh
Richard J. McManus
机构
[1] University of Exeter Medical School,Primary Care Research Group, Institute of Health Services Research
[2] College of Medicine and Health,Nuffield Department of Primary Care Health Sciences
[3] Smeall Building,undefined
[4] St. Luke’s Campus,undefined
[5] University of Oxford,undefined
[6] Radcliffe Primary Care Building,undefined
[7] Radcliffe Observatory Quarter,undefined
来源
关键词
D O I
暂无
中图分类号
学科分类号
摘要
Atrial fibrillation (AF) affects ~3% of the general population and is twice as common with hypertension. Validation protocols for automated sphygmomanometers exclude people with AF, raising concerns over accuracy of hypertension diagnosis or management, using out-of-office blood pressure (BP) monitoring, in the presence of AF. Some devices include algorithms to detect AF; a feature open to misinterpretation as offering accurate BP measurement with AF. We undertook this review to explore accuracy of automated devices, with or without AF detection, for measuring BP. We searched Medline and Embase to October 2018 for studies comparing automated BP measurement devices to a standard mercury sphygmomanometer contemporaneously. Data were extracted by two reviewers. Mean BP differences between devices and mercury were calculated, where not reported and compared; meta-analyses were undertaken where possible. We included 13 studies reporting 14 devices. Mean systolic and diastolic BP differences from mercury ranged from −3.1 to + 6.1/−4.6 to +9.0 mmHg. Considerable heterogeneity existed between devices (I2: 80 to 90%). Devices with AF detection algorithms appeared no more accurate for BP measurement with AF than other devices. A previous review concluded that oscillometric devices are accurate for systolic but not diastolic BP measurement in AF. The present findings do not support that conclusion. Due to heterogeneity between devices, they should be evaluated on individual performance. We found no evidence that devices with AF detection measure BP more accurately in AF than other devices. More home or ambulatory automated BP monitors require validation in populations with AF.
引用
收藏
页码:352 / 364
页数:12
相关论文
共 50 条
  • [1] Accuracy of automated blood pressure measurements in the presence of atrial fibrillation: systematic review and meta-analysis
    Clark, Christopher E.
    McDonagh, Sinead T. J.
    McManus, Richard J.
    JOURNAL OF HUMAN HYPERTENSION, 2019, 33 (05) : 352 - 364
  • [2] Accuracy of automated blood pressure measurement in the presence of atrial fibrillation: systematic review and meta-analysis
    McDonagh, Sinead
    McManus, Richard
    Clark, Christopher
    JOURNAL OF HUMAN HYPERTENSION, 2018, 32 (10) : 706 - 706
  • [3] Automated blood pressure measurement in atrial fibrillation: a systematic review and meta-analysis
    Stergiou, George S.
    Kollias, Anastasios
    Destounis, Antonios
    Tzamouranis, Dimitrios
    JOURNAL OF HYPERTENSION, 2012, 30 (11) : 2074 - 2082
  • [4] Automated blood pressure measurement in atrial fibrillation: a systematic review and meta-analysis
    Cheng, Hao-Min
    Tufanaru, Catalin
    Pearson, Alan
    Chen, Chen-Huan
    JOURNAL OF HYPERTENSION, 2013, 31 (01) : 214 - 215
  • [5] Automated blood pressure measurement in atrial fibrillation: a systematic review and meta-analysis Reply
    Stergiou, George S.
    Kollias, Anastasios
    Destounis, Antonios
    Tzamouranis, Dimitrios
    JOURNAL OF HYPERTENSION, 2013, 31 (01) : 215 - 216
  • [6] Sensitivity and specificity of automated blood pressure devices to detect atrial fibrillation: A systematic review and meta-analysis of diagnostic accuracy
    Tang, Edmond W. L.
    Yip, Benjamin H. K.
    Yu, Chun-Pong
    Wong, Samuel Y. S.
    Lee, Eric K. P.
    FRONTIERS IN CARDIOVASCULAR MEDICINE, 2022, 9
  • [7] ATRIAL FIBRILLATION DETECTION DURING AUTOMATED BLOOD PRESSURE MEASUREMENT: A SYSTEMATIC REVIEW AND META-ANALYSIS
    Kyriakoulis, K.
    Kollias, A.
    Stambolliu, E.
    Stergiou, G. S.
    JOURNAL OF HYPERTENSION, 2018, 36 : E60 - E60
  • [8] AUTOMATED DETECTION OF ATRIAL FIBRILLATION DURING ROUTINE OSCILLOMETRIC MEASUREMENT OF BLOOD PRESSURE: SYSTEMATIC REVIEW AND META-ANALYSIS
    Kyriakoulis, Konstantinos G.
    Kollias, Anastasios
    Chardouvelis, Panagiotis
    Ntousopoulos, Vasileios
    Menti, Ariadni
    Destounis, Antonios
    Stergiou, George S.
    JOURNAL OF HYPERTENSION, 2023, 41 : E20 - E20
  • [9] Predictive validity of automated oscillometric blood pressure monitors for screening atrial fibrillation: a systematic review and meta-analysis
    Park, Seong-Hi
    Jun, Kyung Ja
    Choi, Yun-Kyoung
    EXPERT REVIEW OF MEDICAL DEVICES, 2019, 16 (06) : 503 - 514
  • [10] Measurement reliability of automated oscillometric blood pressure monitor in the elderly with atrial fibrillation: a systematic review and meta-analysis
    Park, Seong-Hi
    Choi, Yun-Kyoung
    BLOOD PRESSURE MONITORING, 2020, 25 (01) : 2 - 12