Investigating the Inconsistency of ITS-90 for SPRTs in the Subrange 0 °C to 419.527 °C

被引:0
|
作者
J. P. Sun
J. T. Zhang
Z. R. Kang
Y. Duan
机构
[1] University of Science and Technology Beijing,
[2] National Institute of Metrology,undefined
[3] Institute of Metrology of Hebei Province,undefined
来源
关键词
Fixed points; Interpolation equation; SPRTs; Subrange inconsistency;
D O I
暂无
中图分类号
学科分类号
摘要
The subrange inconsistency is a significant factor to uncertainty in the standard platinum resistance thermometer (SPRT) subranges of the International Temperature Scale of 1990 (ITS-90). This paper investigated the subrange inconsistency between the water–zinc and water–aluminum subranges. The calibration data for 60 SPRTs from four manufacturers were analyzed, and the result confirms that the coefficient c in the interpolation of ITS-90 is available to determine the subrange inconsistency in this temperature range again. The inconsistency, Δt, can be simply equal to 59.83c.
引用
收藏
页码:1789 / 1794
页数:5
相关论文
共 50 条