Relating volatiles analysis by GC–MS to Oddy test performance for determining the suitability of museum construction materials

被引:0
|
作者
Michael J. Samide
Mary C. Liggett
Jericha Mill
Gregory D. Smith
机构
[1] Butler University,Chemistry Department
[2] Indianapolis Museum of Art at Newfields,Conservation Science Department
来源
关键词
Materials testing; Oddy test; Evolved gas analysis; Chromatography; Volatiles; Pollution; Degradation; Artwork;
D O I
暂无
中图分类号
学科分类号
摘要
Conservators’ decisions regarding the suitability of museum construction materials for use in proximity to artworks still rely heavily on accelerated corrosion tests like the Oddy test despite widespread criticisms. These issues include inconveniently long wait times, sensitivity to only those pollutants capable of tarnishing metals, a general sense of unreliability, exaggerated environmental conditions, and subjectivity in assessing the test’s results. Increasingly, alternative strategies that use instrumental approaches involving volatiles sampling coupled to gas chromatography with mass spectrometry (GC–MS) are being explored as faster, more comprehensive, potentially quantitative, and possible more ‘objective’ means of assessing the dangers of off-gassing from museum construction materials. While many of these characteristics are now well documented, the objectivity of the instrumental result is arguable. While the detection of volatiles and semi-volatiles by GC–MS can confidently yield a list of potential pollutants, “chemical intuition” must be used to predict whether many of the emitted compounds can in fact adversely affect artwork. In this study, evolved gas analysis (EGA) coupled to GC–MS is used to predict the suitability of a small sample set of plastics for use in a museum. The potential impact of volatiles observed in the EGA chromatogram was assessed using chemical reactivity principles and the sparse literature data on the material damages caused by a small group of known pollutants. These same plastics were then tested using the British Museum’s 3-in-1 Oddy test. The prediction based on an educated chemical assessment of the compounds identified through instrumental analysis shows good correlation with pooled results from the Oddy test. In one of the two instances of disagreement, the EGA analysis was actually overly conservative and leaned toward prohibiting or restricting a material that passed the Oddy test. In the other, a material that failed the Oddy test but was passed by instrumental analysis was later shown to contain VOCs that could be considered corrosive. This trial suggests that with practice and experience instrumental approaches may be useful to supplement and perhaps one day supplant traditional accelerated corrosion testing of museum construction materials.
引用
收藏
相关论文
共 4 条
  • [1] Relating volatiles analysis by GC-MS to Oddy test performance for determining the suitability of museum construction materials
    Samide, Michael J.
    Liggett, Mary C.
    Mill, Jericha
    Smith, Gregory D.
    HERITAGE SCIENCE, 2018, 6
  • [2] Evaluating the suitability of museum storage or display materials for the conservation of metal objects: a study on the conformance between the deposited metal film method and the Oddy test
    Jingyi Shen
    Yijia Shen
    Fangyuan Xu
    Xinguang Zhou
    Laiming Wu
    Environmental Science and Pollution Research, 2018, 25 : 35109 - 35129
  • [3] Evaluating the suitability of museum storage or display materials for the conservation of metal objects: a study on the conformance between the deposited metal film method and the Oddy test
    Shen, Jingyi
    Shen, Yijia
    Xu, Fangyuan
    Zhou, Xinguang
    Wu, Laiming
    ENVIRONMENTAL SCIENCE AND POLLUTION RESEARCH, 2018, 25 (35) : 35109 - 35129
  • [4] Instrument Performance Standards: How to Test GC-MS Sensitivity and Performance Quickly and Efficiently for the Analysis of Dioxins and Furans
    Krumwiede, Dirk
    Mehlmann, Heinz
    LC GC NORTH AMERICA, 2016, : 18 - 18