Conditioning and Sampling Issues of EMG Signals in Motion Recognition of Multifunctional Myoelectric Prostheses

被引:0
|
作者
Guanglin Li
Yaonan Li
Long Yu
Yanjuan Geng
机构
[1] Key Lab for Health Informatics of Chinese Academy of Science (CAS),Institute of Biomedical and Health Engineering
[2] Shenzhen Institutes of Advanced Technology,Department of Biology
[3] CAS,undefined
[4] Purdue University,undefined
来源
关键词
Electromyography; Multifunctional myoelectric prosthesis; Signal conditioning and sampling; Limb amputation; Pattern recognition;
D O I
暂无
中图分类号
学科分类号
摘要
Historically, the investigations of electromyography (EMG) pattern recognition-based classification of intentional movements for control of multifunctional prostheses have adopted the filter cut-off frequency and sampling rate that are commonly used in EMG research fields. In practical implementation of a multifunctional prosthesis control, it is desired to have a higher high-pass cut-off frequency to reduce more motion artifacts and to use a lower sampling rate to save the data processing time and memory of the prosthesis controller. However, it remains unclear whether a high high-pass cut-off frequency and a low-sampling rate still preserve sufficient neural control information for accurate classification of movements. In this study, we investigated the effects of high-pass cut-off frequency and sampling rate on accuracy in identifying 11 classes of arm and hand movements in both able-bodied subjects and arm amputees. Compared to a 5-Hz high-pass cut-off frequency, excluding the EMG components below 60 Hz decreased the average accuracy of 0.1% in classifying the 11 movements across able-bodied subjects and increased the average accuracy of 0.1 and 0.4% among the transradial (TR) and shoulder disarticulation (SD) amputees, respectively. Using a 500 Hz instead of a 1-kHz sampling rate, the average classification accuracy only dropped about 2.0% in arm amputees. The combination of sampling rate and high-pass cut-off frequency of 500 and 60 Hz only resulted in about 2.3% decrease in average accuracy for TR amputees and 0.4% decrease for SD amputees in comparison to the generally used values of 1 kHz and 5 Hz. These results suggest that the combination of sampling rate of 500 Hz and high-pass cut-off frequency of 60 Hz should be an optimal selection in EMG recordings for recognition of different arm movements without sacrificing too much of classification accuracy which can also remove most of motion artifacts and power-line interferences for improving the performance of myoelectric prosthesis control.
引用
收藏
页码:1779 / 1787
页数:8
相关论文
共 30 条
  • [1] Conditioning and Sampling Issues of EMG Signals in Motion Recognition of Multifunctional Myoelectric Prostheses
    Li, Guanglin
    Li, Yaonan
    Yu, Long
    Geng, Yanjuan
    ANNALS OF BIOMEDICAL ENGINEERING, 2011, 39 (06) : 1779 - 1787
  • [2] Dexterous motion recognition for myoelectric control of multifunctional transradial prostheses
    Yang, Dapeng
    Gu, Yikun
    Liu, Rongqiang
    Liu, Hong
    ADVANCED ROBOTICS, 2014, 28 (22) : 1533 - 1543
  • [4] EMG Pattern Recognition Control of Multifunctional Prostheses by Transradial Amputees
    Li, Guanglin
    Kuiken, Todd A.
    2009 ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-20, 2009, : 6914 - +
  • [5] Study on Interaction Between Temporal and Spatial Information in Classification of EMG Signals for Myoelectric Prostheses
    Menon, Radhika
    Di Caterina, Gaetano
    Lakany, Heba
    Petropoulakis, Lykourgos
    Conway, Bernard A.
    Soraghan, John J.
    IEEE TRANSACTIONS ON NEURAL SYSTEMS AND REHABILITATION ENGINEERING, 2017, 25 (10) : 1832 - 1842
  • [6] Quantifying Pattern Recognition-Based Myoelectric Control of Multifunctional Transradial Prostheses
    Li, Guanglin
    Schultz, Aimee E.
    Kuiken, Todd A.
    IEEE TRANSACTIONS ON NEURAL SYSTEMS AND REHABILITATION ENGINEERING, 2010, 18 (02) : 185 - 192
  • [7] Intramuscular EMG after targeted muscle reinnervation for pattern recognition control of myoelectric prostheses
    Smith, Lauren H.
    Hargrove, Levi J.
    2013 6TH INTERNATIONAL IEEE/EMBS CONFERENCE ON NEURAL ENGINEERING (NER), 2013, : 1155 - 1158
  • [8] Motion Recognition for Simultaneous Control of Multifunctional Transradial Prostheses
    Jiang, Naifu
    Tian, Lan
    Fang, Peng
    Dai, Yaping
    Li, Guanglin
    2013 35TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC), 2013, : 1603 - 1606
  • [9] Wrist motion pattern recognition system by EMG signals
    Matsumura, Y
    Fukurni, M
    Akamatsu, N
    KNOWLEDGE-BASED INTELLIGENT INFORMATION AND ENGINEERING SYSTEMS, PT 1, PROCEEDINGS, 2005, 3681 : 611 - 617
  • [10] Crosspoint Switching of EMG Signals to Increase Number of Channels for Pattern Recognition Myoelectric Control
    Gajendran, Rudhram
    Tkach, Dennis C.
    Hargrove, Levi J.
    2013 6TH INTERNATIONAL IEEE/EMBS CONFERENCE ON NEURAL ENGINEERING (NER), 2013, : 259 - 262