Optimization of the cooling system in a double multilayer monochromator

被引:0
|
作者
Seungnam Kim
Myeongjin Kim
Chaesun Lee
Jehan Kim
Kwangwoo Kim
机构
[1] Pohang University of Science and Technologies,Pohang Accelerator Laboratory
来源
关键词
Double multilayer monochromator; Double crystal monochromator; DMM; Pohang Light Source;
D O I
暂无
中图分类号
学科分类号
摘要
This paper presents the results of heat transfer and deformation analyses of the first of two monochromator crystals installed in a DMM (double multilayer monochromator) for SAXS (small angle X-ray scattering) beamline at the Pohang Light Source (PLS). The DMM was developed to allow precise rotational and linear motion of the crystal. The two crystals have independent driving systems and can be rotated simultaneously, but the first crystal absorbs most of the heat power, so we used the ANSYS program to calculate the temperature distribution and the deformation in the first crystal and in the crystal holder. When the contact geometry between the crystal and the crystal holder is changed, the surface slope error coming from thermal deformation varies. With these analyses of deformation and slope error, an optimized cooling system was designed for the first crystal holder.
引用
收藏
页码:1724 / 1727
页数:3
相关论文
共 50 条
  • [1] Optimization of the Cooling System in a Double Multilayer Monochromator
    Kim, Seungnam
    Kim, Myeongjin
    Lee, Chaesun
    Kim, Jehan
    Kim, Kwangwoo
    [J]. JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2013, 63 (09) : 1724 - 1727
  • [2] Optimization of a double crystal monochromator
    Jiang, Zheng
    Wang, Eryan
    Song, Ruiqiang
    Guo, Siming
    Wu, Jinjie
    Hou, Dongjie
    An, Zhenghua
    Zhou, Pengyue
    [J]. JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2021, 79 (08) : 697 - 705
  • [3] Laterally graded multilayer double-monochromator
    Macrander, AT
    Als-Nielsen, J
    Liu, C
    Krasnicki, S
    Maj, J
    Mancini, D
    Erdmann, J
    Gaarde, P
    [J]. X-RAY OPTICS DESIGN, PERFORMANCE, AND APPLICATIONS, 1999, 3773 : 100 - 106
  • [4] Optimization of a double crystal monochromator
    Zheng Jiang
    Eryan Wang
    Ruiqiang Song
    Siming Guo
    Jinjie Wu
    Dongjie Hou
    Zhenghua An
    Pengyue Zhou
    [J]. Journal of the Korean Physical Society, 2021, 79 : 697 - 705
  • [5] Compact double multilayer monochromator for SAXS beamline at PLS
    Kim, S. N.
    Kim, M. J.
    Lee, C. S.
    Kim, J. H.
    Koo, C. J.
    [J]. SYNCHROTRON RADIATION INSTRUMENTATION, PTS 1 AND 2, 2007, 879 : 933 - +
  • [6] A DOUBLE MULTILAYER MONOCHROMATOR AT AN ESRF UNDULATOR FOR MICROBEAM EXPERIMENTS
    DESCHAMPS, P
    ENGSTROM, P
    FIEDLER, S
    RIEKEL, C
    WAKATSUKI, S
    HOGHOJ, P
    ZIEGLER, E
    [J]. JOURNAL OF SYNCHROTRON RADIATION, 1995, 2 : 124 - 131
  • [7] Developing a sagittally focusing double-multilayer monochromator
    Wang, Yujie
    Narayanan, Suresh
    Liu, Jinyuan
    Shu, Deming
    Wang, Jin
    [J]. SYNCHROTRON RADIATION INSTRUMENTATION, PTS 1 AND 2, 2007, 879 : 937 - +
  • [8] Development of a Multilayer Monochromator System for the BioCAT Beamline
    Khounsary, Ali
    Chandrashekarappa, Anoop
    Llodra, Anthony
    Antimonov, Mikhail
    Segre, Carlo U.
    Nair, Sudhakar E.
    Irving, Thomas
    [J]. ADVANCES IN X-RAY/EUV OPTICS AND COMPONENTS XI, 2016, 9963
  • [9] Cooling of double crystal monochromator with sagittal focus at SSRF
    WANG Naxiu1
    [J]. Nuclear Science and Techniques, 2010, 21 (02) : 65 - 68
  • [10] Research on the Mechanism of Flow-Induced Vibration in the Cooling System of a Double Crystal Monochromator
    Li, Ao
    Gong, Xuepeng
    Bai, Yang
    Lu, Qipeng
    Li, Shengchi
    Zhang, Wenbo
    Chai, Kewei
    [J]. APPLIED SCIENCES-BASEL, 2024, 14 (07):