Effect of x-ray tube potential on image quality and patient dose for lumbar spine computed radiography examinations

被引:8
|
作者
Brindhaban A. [1 ]
Al Khalifah K. [1 ]
Al Wathiqi G. [2 ]
Al Ostath H. [3 ]
机构
[1] Department of Radiologic Sciences, Kuwait University, Sulaibikhat 90805
[2] Mubarak Al Kabir Hospital, Ministry of Health
[3] Amiri Hospital, Ministry of Health
关键词
Image quality; Lumbar spine; Patient dose; X-ray tube potential;
D O I
10.1007/BF03178721
中图分类号
学科分类号
摘要
A study was conducted to investigate the effect of increasing x-ray tube voltage on patient dose and image quality during computed radiography (CR) examination of the lumbar spine. Images of a pelvis - lumbar spine phantom were obtained at three different tube voltages using two CR systems. The images were evaluated by image scores (IS) using the image quality criteria proposed by the Commission of the European Communities (CEC), by measuring the signal-to-noise ratio and the contrast-to-noise ratio. The entrance surface dose (ESD) to the phantom was also measured using ionisation chambers and the resulting gonad dose (G) and effective dose (E) were calculated. The CEC recommended tube voltages can be exceeded with CR systems for lumbar spine radiography. The ESD, G and E decreased by 25% - 50% with both CR systems when tube voltage was increased. Although the decrease in IS was statistically significant (P<0.05) the images obtained with the 3 tube voltages were diagnostically acceptable. The balance between image quality and patient dose can be optimised by using higher x-ray tube voltages for lumbar spine CR examinations. Copyright © 2005 ACPSEM/EA.
引用
收藏
页码:216 / 222
页数:6
相关论文
共 50 条
  • [1] Comparison of image quality and patient dose for chest x-ray examinations on conventional and low cost computed radiography systems
    Muhogora, W. E.
    Trianni, A.
    Toso, F.
    Devetti, A.
    Padovani, R.
    Msaki, P.
    Kazema, R.
    RADIOGRAPHY, 2012, 18 (04) : 275 - 278
  • [3] PATIENT DOSE AND IMAGE QUALITY IN 5 STANDARD X-RAY EXAMINATIONS
    HAVUKAINEN, R
    PIRINEN, M
    MEDICAL PHYSICS, 1993, 20 (03) : 813 - 817
  • [4] Effect of varying X-ray tube voltage and additional filtration on image quality and patient dose in digital radiography system
    Aksit, E.
    Yalcin, A.
    Olgar, T.
    APPLIED RADIATION AND ISOTOPES, 2023, 199
  • [5] THE INFLUENCE OF TUBE FILTRATION AND POTENTIAL ON PATIENT DOSE DURING X-RAY EXAMINATIONS
    SHRIMPTON, PC
    JONES, DG
    WALL, BF
    PHYSICS IN MEDICINE AND BIOLOGY, 1988, 33 (10): : 1205 - 1212
  • [6] DOSE AND IMAGE QUALITY CALCULATIONS FOR CHEST X-RAY EXAMINATIONS
    KOBLINGE.L
    ZARAND, P
    PHYSICS IN MEDICINE AND BIOLOGY, 1972, 17 (05): : 712 - &
  • [7] Effect of Tube Setting on Image Quality in Industrial X-ray Computed Tomography
    Sun, Zhaoying
    Li, Lei
    Xi, Xiaoqi
    Han, Yu
    Yan, Bin
    Rong, Lihui
    Zhou, Sangsang
    THREE-DIMENSIONAL IMAGE ACQUISITION AND DISPLAY TECHNOLOGY AND APPLICATIONS, 2018, 10845
  • [8] Comparison of patient radiation dose from chest and lumbar spine X-ray examinations in 10 hospitals in Ghana
    Ofori, E. K.
    Antwi, W. K.
    Arthur, L.
    Duah, H.
    RADIATION PROTECTION DOSIMETRY, 2012, 149 (04) : 424 - 430
  • [9] The influence of the characteristic curve on the clinical image quality and patient absorbed dose in lumbar spine radiography
    Tingberg, A
    Herrmann, C
    Lanhede, B
    Almén, A
    Mattsson, S
    Panzer, W
    Besjakov, J
    Månsson, LG
    Kheddache, S
    Zankl, M
    MEDICAL IMAGING 2001: IMAGE PERCEPTION AND PERFORMANCE, 2001, 4324 : 8 - 17
  • [10] X-RAY MAMMOGRAPHY - IMAGE QUALITY AND PATIENT DOSE
    MASTERSON, ME
    MEDICAL PHYSICS, 1985, 12 (04) : 555 - 555