Estimation of mechanical heart valve cavitation in a pneumatic ventricular assist device
被引:0
|作者:
Hwansung Lee
Eiki Akagawa
Akihiko Homma
Tomonori Tsukiya
Eisuke Tatsumi
Yoshiyuki Taenaka
机构:
[1] National Cardiovascular Center,Department of Artificial Organs, Research Institute
来源:
关键词:
Pneumatic ventricular assist device;
Mechanical heart valve;
Cavitation;
Driving pressure;
D O I:
暂无
中图分类号:
学科分类号:
摘要:
In this study, we investigated the possibility of estimating the mechanical heart valve (MHV) cavitation intensity using the slope of the driving pressure (DP) just before valve closure in a pneumatic ventricular assist device. We installed a 23-mm Medtronic Hall valve at the inlet of our pneumatic ventricular assist device (VAD). Tests were conducted under physiologic pressures at heart rates ranging from 60 to 90 beats/min and cardiac outputs ranging from 4.5 to 6.7 l/min. The valve-closing velocity was measured with a CCD laster displacement sensor, and the images of MHV cavitation were recorded using a high-speed video camera. The cavitation cycle time (equal to the observed duration of the cavitation bubbles) was used as the MHV cavitation intensity. The valve-closing velocity increased as the heart rate increased. Most of the cavitation bubbles were observed near the valve stop, and the cavitation intensity increased as the heart rate increased. The slope of the DP at 20 ms before valve closure was used as an index of the cavitation intensity. There were differences in the slope of the DP between low and high heart rates, but the slope of the DP had a tendency to linearly increase with increasing valve-closing velocity.
引用
收藏
页码:181 / 185
页数:4
相关论文