Estimation of mechanical heart valve cavitation in a pneumatic ventricular assist device

被引:0
|
作者
Hwansung Lee
Eiki Akagawa
Akihiko Homma
Tomonori Tsukiya
Eisuke Tatsumi
Yoshiyuki Taenaka
机构
[1] National Cardiovascular Center,Department of Artificial Organs, Research Institute
来源
关键词
Pneumatic ventricular assist device; Mechanical heart valve; Cavitation; Driving pressure;
D O I
暂无
中图分类号
学科分类号
摘要
In this study, we investigated the possibility of estimating the mechanical heart valve (MHV) cavitation intensity using the slope of the driving pressure (DP) just before valve closure in a pneumatic ventricular assist device. We installed a 23-mm Medtronic Hall valve at the inlet of our pneumatic ventricular assist device (VAD). Tests were conducted under physiologic pressures at heart rates ranging from 60 to 90 beats/min and cardiac outputs ranging from 4.5 to 6.7 l/min. The valve-closing velocity was measured with a CCD laster displacement sensor, and the images of MHV cavitation were recorded using a high-speed video camera. The cavitation cycle time (equal to the observed duration of the cavitation bubbles) was used as the MHV cavitation intensity. The valve-closing velocity increased as the heart rate increased. Most of the cavitation bubbles were observed near the valve stop, and the cavitation intensity increased as the heart rate increased. The slope of the DP at 20 ms before valve closure was used as an index of the cavitation intensity. There were differences in the slope of the DP between low and high heart rates, but the slope of the DP had a tendency to linearly increase with increasing valve-closing velocity.
引用
收藏
页码:181 / 185
页数:4
相关论文
共 50 条
  • [1] Estimation of mechanical heart valve cavitation in a pneumatic ventricular assist device
    Lee, Hwansung
    Akagawa, Eiki
    Hornma, Akihiko
    Tsukiya, Tomonori
    Tatsumi, Eisuke
    Taenaka, Yoshiyuki
    [J]. JOURNAL OF ARTIFICIAL ORGANS, 2007, 10 (03) : 181 - 185
  • [2] Characteristics of mechanical heart valve cavitation in a pneumatic ventricular assist device
    Lee, Hwansung
    Taenaka, Yoshiyuki
    [J]. ARTIFICIAL ORGANS, 2008, 32 (06) : 453 - 460
  • [3] Effect of Systolic duration on mechanical heart valve cavitation in a pneumatic ventricular assist device: Using a monoleaflet valve
    Lee, Hwansung
    Tatsumi, Eisuke
    Taenaka, Yoshiyuki
    [J]. ASAIO JOURNAL, 2008, 54 (01) : 25 - 30
  • [4] Effects of the Driving Condition of a Pneumatic Ventricular Assist Device on the Cavitation Intensity of the Inlet and Outlet Mechanical Heart Valves
    Lee, Hwansung
    Tatsumi, Eisuke
    Taenaka, Yoshiyuki
    [J]. ASAIO JOURNAL, 2009, 55 (04) : 328 - 334
  • [5] Flow Visualization of A Monoleaflet and Bileaflet Mechanical Heart Valve in A Pneumatic Ventricular Assist Device Using A PIV System
    Lee, Hwansung
    Tatsumi, Eisuke
    Taenaka, Yoshiyuki
    [J]. ASAIO JOURNAL, 2010, 56 (03) : 186 - 193
  • [6] Mechanical heart valve performance in a pulsatile pediatric ventricular assist device
    Zapanta, Conrad M.
    Dourte, Leann M.
    Doxtater, Bradley J.
    Lukic, Branka
    Weiss, William J.
    [J]. ASAIO JOURNAL, 2007, 53 (01) : 87 - 96
  • [7] Hydrodynamic characteristics of the Edwards MIRA bileaflet valve in a pneumatic ventricular assist device
    Lee, Hwansung
    Taenaka, Yoshiyuki
    [J]. ASAIO JOURNAL, 2007, 53 (04) : 397 - 402
  • [8] Experimental Study on the Reynolds and Viscous Shear Stress of Bileaflet Mechanical Heart Valves in a Pneumatic Ventricular Assist Device
    Lee, Hwansung
    Tatsumi, Eisuke
    Taenaka, Yoshiyuki
    [J]. ASAIO JOURNAL, 2009, 55 (04) : 348 - 354
  • [9] Pneumatic Pulsatile Ventricular Assist Device as a Bridge to Heart Transplantation in Pediatric Patients
    Amodeo, Antonio
    Brancaccio, Gianluca
    Michielon, Guido
    Filippelli, Sergio
    Ricci, Zaccaria
    Morelli, Stefano
    Gagliardi, Maria Giulia
    Iacobelli, Roberta
    Pongiglione, Giacomo
    Di Donato, Roberto M.
    [J]. ARTIFICIAL ORGANS, 2010, 34 (11) : 1017 - 1022
  • [10] Effect of the diastolic and systolic duration on valve cavitation in a pediatric pulsatile ventricular assist device
    Lukic, Branka
    Zapanta, Conrad M.
    Griffith, Kimberly A.
    Weiss, William J.
    [J]. ASAIO JOURNAL, 2005, 51 (05) : 546 - 550