Ir-192 Calibration in Air with Farmer Chamber for HDR Brachytherapy

被引:0
|
作者
Liyun Chang
Sheng-Yow Ho
Tsair-Fwu Lee
Hueisch-Jy Ding
Pang-Yu Chen
机构
[1] I-Shou University,Department of Medical Imaging and Radiological Sciences
[2] Chang Jung Christian University,Department of Nursing
[3] Chi Mei Medical Center,Department of Radiation Oncology
[4] National Kaohsiung University of Applied Sciences,Medical Physics and Informatics Laboratory, Department of Electronics Engineering
[5] Sinlau Christian Hospital,Department of Radiation Oncology
关键词
Ir-192 calibration; In-air calibration; Farmer-type ion chamber; High dose rate (HDR);
D O I
暂无
中图分类号
学科分类号
摘要
A comprehensive review for the in-air calibration of an Ir-192 high-dose-rate brachytherapy source in terms of air kerma strength (AKS) and reference air kerma rate (RAKR) using the Farmer chamber was conducted. The reviewed calibration methods include the National Physical Laboratory (NPL) calibration standard in the UK, the 7-distance technique with the standard calibration of the National Institute of Standards and Technology and Accredited Dosimetry Calibration Laboratory in the US, the calibration conducted in Australia following recommendations of the International Atomic Energy Agency with the chamber primarily calibrated by the Australian Radiation Protection and Nuclear Safety Agency, the calibration conducted in India following the Deutsche Gesellschaft fur Medizinische Physik recommendation, and the convenient empirical method used in Taiwan. The calibrated AKS (or RAKR) and uncertainty obtained using Farmer chambers are similar to those obtained using well chambers. All reported differences (between measurements obtained using Farmer and well chambers, respectively) and uncertainties (k = 2) were generally less than 1.5 and 2.5 %, respectively. The standard uncertainty of the NPL calibration is approximately half that of all the other proposed approaches, and may become the gold standard calibration procedure. Almost all techniques follow the 7-distance technique basis; however, the services at NPL can calibrate the source with lower uncertainty. Users can calibrate the Ir-192 source more conveniently using the empirical method with only one source-chamber distance.
引用
收藏
页码:145 / 152
页数:7
相关论文
共 50 条
  • [1] Ir-192 Calibration in Air with Farmer Chamber for HDR Brachytherapy
    Chang, Liyun
    Ho, Sheng-Yow
    Lee, Tsair-Fwu
    Ding, Hueisch-Jy
    Chen, Pang-Yu
    [J]. JOURNAL OF MEDICAL AND BIOLOGICAL ENGINEERING, 2016, 36 (02) : 145 - 152
  • [2] Absolute calibration of HDR IR-192 brachytherapy sources
    Sander, S.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2006, 81 : S178 - S179
  • [3] CALIBRATION OF BRACHYTHERAPY IR-192 SOURCES
    COBB, PD
    CHEN, TS
    KASE, KR
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1981, 7 (02): : 259 - 262
  • [4] CALIBRATION OF A MICROSELECTRON HDR IR-192 SOURCE
    FLYNN, A
    WORKMAN, G
    [J]. BRITISH JOURNAL OF RADIOLOGY, 1991, 64 (764): : 734 - 739
  • [5] A DOSIMETRIC STUDY OF THE MICROSELECTRON HDR IR-192 BRACHYTHERAPY SOURCE
    Granero, D.
    Rivard, M.
    Ballester, F.
    Perez-Calatayud, J.
    Vijande, J.
    Otal, A.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2010, 96 : S479 - S479
  • [6] A Dosimetric Study of the MicroSelectron HDR Ir-192 Brachytherapy Source
    Granero, D.
    Rivard, M. J.
    Ballester, F.
    Perez-Calatayud, J.
    Vijande, J.
    Otal, A.
    [J]. MEDICAL PHYSICS, 2010, 37 (06)
  • [7] Dosimetry for Ir-192 HDR Brachytherapy - Present Status and Future Direction
    DeWerd, L.
    Seuntjens, J.
    Rivard, M.
    McEwen, M.
    [J]. MEDICAL PHYSICS, 2012, 39 (06) : 3884 - 3885
  • [8] Monte Carlo study of a bunker used for HDR Ir-192 brachytherapy
    Ballester, F
    Granero, D
    Pérez-Calatayud, J
    Puchades, V
    Casal, E
    Crispin, V
    Cases, R
    Agramunt, S
    [J]. RADIOTHERAPY AND ONCOLOGY, 2004, 73 : S242 - S242
  • [9] Adjuvant HDR-Brachytherapy with Ir-192 in the Treatment of recurrent Keloids
    Jiang, P.
    Bockelmann, G.
    Niehoff, P.
    Donner, H.
    Siebert, F. -A
    Druecke, D.
    Dunst, J.
    Kimmig, B.
    [J]. STRAHLENTHERAPIE UND ONKOLOGIE, 2013, 189 : 51 - 52
  • [10] Quality Control for Cervical Ca treatments on HDR Brachytherapy with Ir-192
    Alvarino Bettin, G.
    Cogollo Pitalua, R.
    Paez Meza, M.
    [J]. REVISTA MEXICANA DE FISICA, 2013, 59 (01) : 38 - 44