Cone beam computed tomography (CBCT) is now widely used to image radiotherapy patients prior to treatment for the purpose of accurate patient setup. However each CBCT image delivered to a patient increases the total radiation dose that they receive. The measurement of the dose delivered from the CBCT images is not readily performed in the clinic. In this study, we have used commercially available optically stimulated luminescence (OSLD) dosimeters to measure the dose delivered by the Varian OBI on a radiotherapy linear accelerator. Calibration of the OSLDs was achieved by using a therapeutic X-ray unit. The dose delivered by a head CBCT scan was found to be 3.2 ± 0.3 mGy which is similar in magnitude to the dose of a head computed tomography (CT) scan. The results of this study suggest that the radiation hazard associated with CBCT is of a similar nature to that of conventional CT scans. We have also demonstrated that the OSLDs are suitable for these low X-ray dose measurements.
引用
收藏
页码:665 / 669
页数:4
相关论文
共 50 条
  • [1] Radiation dose measurements of an on-board imager X-ray unit using optically-stimulated luminescence dosimeters
    Smith, Leon
    Haque, Mamoon
    Morales, Johnny
    Hill, Robin
    AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE, 2015, 38 (04) : 665 - 669
  • [2] DESIGN OF INTERROGATION PROTOCOLS FOR RADIATION DOSE MEASUREMENTS USING OPTICALLY-STIMULATED LUMINESCENT DOSIMETERS
    Abraham, Sara A.
    Kearfott, Kimberlee J.
    Jawad, Ali H.
    Boria, Andrew J.
    Buth, Tobias J.
    Dawson, Alexander S.
    Eng, Sheldon C.
    Frank, Samuel J.
    Green, Crystal A.
    Jacobs, Mitchell L.
    Liu, Kevin
    Miklos, Joseph A.
    Hien Nguyen
    Rafique, Muhammad
    Rucinski, Blake D.
    Smith, Travis
    Tan, Yanliang
    HEALTH PHYSICS, 2017, 112 (03): : 237 - 245
  • [3] Radiation dose measurements in a dental orthopantogram unit using indigenously developed optically stimulated luminescence dosimeters
    Kumar, Pradip
    Sharma, Sunil Dutt
    Chandola, Rakesh Mohan
    Mishra, Devesh Ramdhar
    Dhabekar, Bhushan
    Rawat, Narender Singh
    Kadam, Sonal
    Agrawal, Sadhana
    LUMINESCENCE, 2019, 34 (04) : 444 - 449
  • [4] Optically-Stimulated Luminescent Dosimeters for Monitoring Pacemaker Dose in Radiation Therapy
    Apicello, L.
    Riegel, A.
    Jamshidi, A.
    MEDICAL PHYSICS, 2015, 42 (06) : 3470 - 3470
  • [5] A methodology for on-board CBCT imaging dose using optically stimulated luminescence detectors
    Mail, Noor
    Yusuf, Muhammad
    Alothmany, Nazeeh
    Kinsara, A. Abdulrahman
    Abdulkhaliq, Fahad
    Ghamdi, Suliman M.
    Saoudi, Abdelhamid
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2016, 17 (05): : 482 - 499
  • [6] Measurement of radiation dose with BeO dosimeters using optically stimulated luminescence technique in radiotherapy
    Sahin, Serdar
    Tanir, A. Gunes
    Meric, Niyazi
    Aydinkarahaliloglu, Ercan
    APPLIED RADIATION AND ISOTOPES, 2015, 103 : 31 - 36
  • [7] Dosimetric measurement of testicular dose for colorectal cancer using optically-stimulated luminescent dosimeters in radiotherapy
    Lu, C. C.
    Wang, F. N.
    Lin, H. H.
    Hsu, C. H.
    Lin, J. P.
    Lai, L. H.
    RADIATION PHYSICS AND CHEMISTRY, 2020, 172
  • [8] Comparison of the verification performance and radiation dose between ExacTrac x-ray system and On-Board Imager-A phantom study
    Wing-cheung, Vincent
    Ho, Yan-yee
    Tang, Yu-sum
    Lam, Po-wa
    Yeung, Ho-kwan
    Lee, Shara Wee-yee
    MEDICAL DOSIMETRY, 2019, 44 (01) : 15 - 19
  • [9] Dosimetric Assessment of Routine X-Ray Examination at Selected Health Clinics in Perak Using Commercialized Optically-Stimulated Luminescence Dosimeter (OSLD)
    Saidin, M. T.
    Rahman, A. A.
    Harun, H. H.
    Radzi, Y. M.
    Anam, C.
    Kayun, Z.
    ATOM INDONESIA, 2021, 47 (03) : 213 - 218
  • [10] Thermoluminescent dosimeters for low dose X-ray measurements
    Del Sol Fernandez, S.
    Garcia-Salcedo, R.
    Sanchez-Guzman, D.
    Ramirez-Rodriguez, G.
    Gaona, E.
    de Leon-Alfaro, M. A.
    Rivera-Montalvo, T.
    APPLIED RADIATION AND ISOTOPES, 2016, 107 : 340 - 345