Patient-specific finite element analysis of popliteal stenting

被引:0
|
作者
Michele Conti
Michele Marconi
Giulia Campanile
Alessandro Reali
Daniele Adami
Raffaella Berchiolli
Ferdinando Auricchio
机构
[1] Pavia University,Department of Civil Engineering and Architecture (DICAr)
[2] IRCCS Cisanello,Vascular Unit
[3] Technische Universität München,Institute for Advanced Study
来源
Meccanica | 2017年 / 52卷
关键词
Finite element analysis (FEA); Biomechanics; Popliteal stenting; Patient-specific; Nitinol;
D O I
暂无
中图分类号
学科分类号
摘要
Nitinol self-expanding stents are used for the endovascular management of peripheral artery diseases of the popliteal artery, which is located behind the knee joint. Unfortunately, the complex kinematics of the artery during the leg flexion leads to severe loading conditions, favouring the mechanical failure of the stent, calling for a specific biomechanical analysis. For this reason, in the present study we reconstruct by medical image analysis the patient-specific popliteal kinematics during leg flexion, which is subsequently exploited to compute the mechanical response of a stent model, virtually implanted in the artery by structural finite element analysis (FEA). The medical image analysis indicates a non-uniform configuration change of the artery during the leg flexion, leading to an increase of the vessel curvature above the knee. The computed mechanical response of the stent reflects the non-uniform configuration change of the artery as after the flexion the average stress is higher in the part of the stent located above the knee. Although the proposed analysis is limited to a case-study, it shows the capability of patient-specific FEA simulations to compute the mechanical response of a stent model subjected to the complex and severe loading conditions of the popliteal artery during leg flexion.
引用
收藏
页码:633 / 644
页数:11
相关论文
共 50 条
  • [1] Patient-specific finite element analysis of popliteal stenting
    Conti, Michele
    Marconi, Michele
    Campanile, Giulia
    Reali, Alessandro
    Adami, Daniele
    Berchiolli, Raffaella
    Auricchio, Ferdinando
    [J]. MECCANICA, 2017, 52 (03) : 633 - 644
  • [2] PATIENT-SPECIFIC FINITE ELEMENT ANALYSIS OF CAROTID ARTERY STENTING
    Conti, Michele
    Auricchio, Ferdinando
    De Santis, Gianluca
    De Beule, Matthieu
    Verhegghe, Benedict
    [J]. PROCEEDINGS OF THE ASME SUMMER BIOENGINEERING CONFERENCE, 2010, 2010, : 561 - 562
  • [3] Patient-specific Finite Element Model of Coronary Artery Stenting
    Razaghi, Reza
    Karimi, Alireza
    Taheri, Ramezan Ali
    [J]. CURRENT PHARMACEUTICAL DESIGN, 2018, 24 (37) : 4492 - 4502
  • [4] Patient-specific finite element analysis of carotid artery stenting: a focus on vessel modeling
    Auricchio, F.
    Conti, M.
    Ferrara, A.
    Morganti, S.
    Reali, A.
    [J]. INTERNATIONAL JOURNAL FOR NUMERICAL METHODS IN BIOMEDICAL ENGINEERING, 2013, 29 (06) : 645 - 664
  • [5] Carotid artery stenting simulation: From patient-specific images to finite element analysis
    Auricchio, F.
    Conti, M.
    De Beule, M.
    De Santis, G.
    Verhegghe, B.
    [J]. MEDICAL ENGINEERING & PHYSICS, 2011, 33 (03) : 281 - 289
  • [6] Prediction of stenting related adverse events through patient-specific finite element modelling
    Caimi, Alessandro
    Sturla, Francesco
    Pluchinotta, Francesca Romana
    Giugno, Luca
    Secchi, Francesco
    Votta, Emiliano
    Carminati, Mario
    Redaelli, Alberto
    [J]. JOURNAL OF BIOMECHANICS, 2018, 79 : 135 - 146
  • [7] Patient-specific finite element analysis of ascending aorta aneurysms
    Martin, Caitlin
    Sun, Wei
    Elefteriades, John
    [J]. AMERICAN JOURNAL OF PHYSIOLOGY-HEART AND CIRCULATORY PHYSIOLOGY, 2015, 308 (10): : H1306 - H1316
  • [8] Impact of leg bending in the patient-specific computational fluid dynamics of popliteal stenting
    A. Ferrarini
    A. Finotello
    G. Salsano
    F. Auricchio
    D. Palombo
    G. Spinella
    B. Pane
    M. Conti
    [J]. Acta Mechanica Sinica, 2021, 37 : 279 - 291
  • [9] Patient-specific finite element modeling of bones
    Poelert, Sander
    Valstar, Edward
    Weinans, Harrie
    Zadpoor, Amir A
    [J]. Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine, 2012, 227 (04) : 464 - 478
  • [10] Impact of leg bending in the patient-specific computational fluid dynamics of popliteal stenting
    Ferrarini, A.
    Finotello, A.
    Salsano, G.
    Auricchio, F.
    Palombo, D.
    Spinella, G.
    Pane, B.
    Conti, M.